Fig. 15

Effect of LED temperature on light flux (warming of CFHT dome at dawn): a) variable exposure (1s < Δt < 8s); b) constant exposure (Δt = 8s): two independent estimators ⟨ αk ⟩≈ 16 000 adu (black points) and ⟨ Ψσδ ⟩≈ 100 adu (red points) agree within 0.8 × 10-4 rms. Deviation from linearity is 1.8 × 10-4 rms. The precision on ⟨ Ψσδ ⟩ is ≈0.008 adu, i.e., 0.6 × 10-6. Error bars cover gain spread before averaging. The two other estimators ⟨ Ψσβ ⟩ and ⟨ Ψσγ ⟩ (green and blue) are more sensitive to LED jitter.