Pre-launch status of the Planck mission
Planck pre-launch status: Calibration of the Low Frequency Instrument flight model radiometers
F. Villa1
- L. Terenzi1 - M. Sandri1
- P. Meinhold2 - T. Poutanen3,4,5
- P. Battaglia6 - C. Franceschet7
- N. Hughes8 - M. Laaninen8
- P. Lapolla6 - M. Bersanelli7
- R. C. Butler1 - F. Cuttaia1
- O. D'Arcangelo9 - M. Frailis10
- E. Franceschi1 - S. Galeotta10
- A. Gregorio11 - R. Leonardi2
- S. R. Lowe12 - N. Mandolesi1
- M. Maris10 - L. Mendes13
- A. Mennella7 - G. Morgante1
- L. Stringhetti1,
- M. Tomasi7 - L. Valenziano1
- A. Zacchei10 - A. Zonca14
- B. Aja15 - E. Artal15
- M. Balasini6 - T. Bernardino16
- E. Blackhurst12 - L. Boschini6
- B. Cappellini14 - F. Cavaliere7
- A. Colin16 - F. Colombo6
- R. J. Davis12 - L. De La Fuente15
- J. Edgeley12 - T. Gaier17
- A. Galtress12 - R. Hoyland18
- P. Jukkala8 - D. Kettle12
- V.-H. Kilpia8 - C. R. Lawrence16
- D. Lawson12 - J. P. Leahy12
- P. Leutenegger6 - S. Levin16
- D. Maino7 - M. Malaspina1
- A. Mediavilla15 - M. Miccolis6
- L. Pagan6 - J. P. Pascual15
- F. Pasian10 - M. Pecora6
- M. Pospieszalski19 - N. Roddis12
- M. J. Salmon16 - M. Seiffert17
- R. Silvestri6 - A. Simonetto9
- P. Sjoman8 - C. Sozzi9
- J. Tuovinen20 - J. Varis 20
- A. Wilkinson12 - F. Winder12
1 - INAF - Istituto di Astrofisica Spaziale e Fisica Cosmica, via P.
Gobetti, 101, 40129 Bologna, Italy
2 - Department of Physics, University of California, Santa Barbara,
93106-9530, USA
3 - University of Helsinki, Department of Physics, PO Box 64, 00014
Helsinki, Finland
4 - Helsinki Institute of Physics, PO Box 64, 00014 Helsinki, Finland
5 - Metsähovi Radio Observatory, Helsinki University of Technology,
Metsähovintie 114, 02540 Kylmälä, Finland
6 - Thales Alenia Space Italia S.p.A., S.S. Padana Superiore 290, 20090
Vimodrone (MI), Milano, Italy
7 - Università degli Studi di Milano, via Celoria 16, 20133 Milano,
Italy
8 - DA-Design Oy. (aka Ylinen Electronics), Keskuskatu 29, 31600
Jokioinen, Finland
9 - IFP-CNR, via Cozzi 53, 20125 Milano, Italy
10 - INAF - Osservatorio Astronomico di Trieste, 11 via Tiepolo, 34143
Trieste, Italy
11 - University of Trieste, Department of Physics, 2 via Valerio, 34127
Trieste, Italy
12 - Jodrell Bank Centre for Astrophysics, Alan Turing Building, The
University of Manchester, Manchester, M13 9PL, UK
13 - Planck Science Office, European Space Agency, ESAC, PO box 78,
28691 Villanueva de la Cañada, Madrid, Spain
14 - INAF - Istituto di Astrofisica Spaziale e Fisica Cosmica, via E.
Bassini 15, 20133 Milano, Italy
15 - Universidad de Cantabria, Dep. De Ingenieria de Comunicaciones.
Av. Los Castros s/n, 39005, Santander-Spain
16 - Instituto de Fisica de Cantabria (CSIC-UC), Av. Los Castros s/n,
39005 Santander, Spain
17 - Jet Propulsion Laboratory, 4800 Oak Grove Drive, Pasadena,
California 91109, USA
18 - Instituto de Astrofísica de Canarias C/, Via Lactea s/n, 38205 La
Laguna (Tenerife), Spain
19 - National Radio Astronomy Observatory, Stone hall, University of
Virginia, 520 Edgemont road, Charlottesville, USA
20 - MilliLab, VTT Technical Research Centre of Finland, Tietotie 3,
Otaniemi, Espoo, Finland
Received 9 July 2009 / Accepted 3 May 2010
Abstract
The Low Frequency Instrument (LFI) on-board the ESA Planck
satellite carries eleven radiometer subsystems, called radiometer chain
assemblies (RCAs), each composed of a pair of pseudo-correlation
receivers. We describe the on-ground calibration campaign performed to
qualify the flight model RCAs and to measure their pre-launch
performances.
Each RCA was calibrated in a dedicated flight-like cryogenic
environment with the radiometer front-end cooled to 20 K and
the back-end at 300 K, and with an external input load cooled
to 4 K. A matched load simulating a blackbody at different
temperatures was placed in front of the sky horn to derive basic
radiometer properties such as noise temperature, gain, and noise
performance, e.g. 1/f noise. The spectral response
of each detector was measured as was their susceptibility to thermal
variation.
All eleven LFI RCAs were calibrated. Instrumental parameters measured
in these tests, such as noise temperature, bandwidth, radiometer
isolation, and linearity, provide essential inputs to the Planck-LFI
data analysis.
Key words: cosmic microwave background - space vehicles: instruments - instrumentation: detectors - techniques: miscellaneous
1 Introduction
The Planck mission
has been developed to provide a deep, full-sky image of the cosmic
microwave background (CMB) in both temperature and polarization. Planck
incorporates an unprecedented combination of sensitivity, angular
resolution and spectral range - spanning from centimeter to
sub-millimeter wavelengths - by integrating two complementary cryogenic
instruments in the focal plane of the Planck
telescope. The Low Frequency Instrument (LFI) covers the region below
the CMB blackbody peak in three frequency bands centered
at 30, 44 and 70 GHz. The spectral range of the LFI
is also suitable for a wealth of galactic and extragalactic
astrophysics. The LFI maps will address studies of diffuse Galactic
free-free and synchrotron emission, emission from spinning dust grains,
and discrete Galactic radio sources. Extragalactic radio sources will
also be observed, particularly those with flat or strongly inverted
spectra, peaking at mm wavelengths. Furthermore, the Planck
scanning strategy will also allow monitoring of radio source
variability on a variety of time scales. While these are highly
interesting astrophysical objectives, the LFI design and calibration is
driven by the main Planck scientific scope, i.e.,
CMB science.
![]() |
Figure 1: Scheme of the RCA and its flight-like thermal interfaces. The FEM is at 20 K, while the BEM is at 315 K. At 30 and 44 GHz the thermal interface attached to the third and coldest V-groove (VG-3) is controlled at a temperature near 60 K, while for the 70 GHz RCA the VG-3 interface was not controlled in temperature. Both loads - the sky load and the reference load - are controlled in a temperature of approximately 4 K to 35 K. |
Open with DEXTER |
The LFI is an array of cryogenic radiometers based on indium phospide
(InP) cryogenic HEMT low noise amplifiers (Bersanelli
et al. 2010). The array is composed of 22
pseudo-correlation radiometers mounted in eleven independent radiometer
units called ``radiometer chain assemblies'' (RCAs), two centered at
30 GHz, three at 44 GHz and six at 70 GHz. To optimize the
sensitivity and minimize the power dissipation in the front end, each
RCA is split into a front-end module (FEM), cooled to 20 K,
and a back-end module (BEM), operating at 300 K, connected by
a set of composite waveguides.
Accurate calibration is mandatory for optimal operation of the instrument during the full-sky survey and for measuring parameters that are essential for the Planck data analysis.
The LFI calibration strategy has been based on a complementary approach that includes both pre-launch and post-launch activities. On-ground measurements were performed at all-unit and sub-unit levels, both for qualification and performance verification. Each single FEM and BEM as well as the passive components (feed horn, orthomode transducers, waveguides, 4 K reference loads) were tested in a stand-alone configuration before they were integrated into the RCA units.
The final scientific calibration of the LFI was carried out at
two different integration levels, depending on the measured parameter.
First each RCA was tested independently in dedicated cryofacilities,
which are capable of reaching a temperature of 4 K at the external input loads. These
conditions were necessary for an accurate measurement of key parameters
such as system noise temperature, bandwidth, radiometer isolation, and
linearity. Subsequently, the eleven RCAs were integrated into the full
LFI instrument (the so-called radiometer array assembly, RAA) and
tested as a complete instrument system in a large cryofacility, with
highly stable input loads cooled down to 20 K.
We reports on the calibration campaign of the RCAs, while Mennella et al. (2010) report on the RAA calibration. The parameters derived here are crucial for the Planck-LFI scientific analysis. Noise temperatures, bandwidths, and radiometer isolation provide essential information to construct an adequate noise model, which is needed as an input to the map-making process. Any non-linearity of the instrument response must be accurately measured, because corrections may be needed in the data analysis, particularly for observations of strong sources such as planets (crucial for in-flight beam reconstruction) and the Galactic plane. As part of the RCA testing, we also performed an end-to-end measurement of the bandshape inside the cryofacility. Finally, for comparison and as a consistency check, the RCA test plan also included measurement of parameters whose primary calibration relies on the RAA test campaign, such as optimal radiometer bias (tuning), 1/f noise (knee frequency and slope), gain, and thermal susceptibility.
The 30 and 44 GHz RCAs were integrated and tested in Thales Alenia Space Italia (TAS-I), formerly Laben, from the beginning of January 2006 to end of May 2006 using a dedicated cryofacility (Terenzi et al. 2009b) to reproduce flight-like thermal interfaces and input loads. The 70 GHz RCAs were integrated and calibrated in Yilinen Electronics (Finland) from the end of April 2005 to mid February 2006 with a similar cryofacility, but with simplified thermal interfaces (Terenzi et al. 2009b). The differences between the two cryofacilities resulted in slightly different test procedures, because the temperatures were not controlled in the same way.
Section 2 of the paper describes the concept of RCA calibration, while Sect. 3 illustrates the two cryofacilities. In Sect. 4 we describe each test, the methods used in the analysis and the results. The conclusions are given in Sect. 5.
2 Main concepts and calibration logic
2.1 Radiometer chain assembly description
A diagram of an RCA is shown in Fig. 1.
A corrugated feed horn (Villa
et al. 2009), which collects the radiation from the
telescope, ,
is connected to an ortho-mode transducer (D'Arcangelo
et al. 2009b), which divides the signal into two
orthogonal polarizations, namely ``M'' (main) and ``S'' (side)
branches. The OMT is connected to the front-end section (Varis
et al. 2009; Davis et al. 2009), in
which each polarization of the sky signal is mixed with the signal from
a stable reference load,
,
via a hybrid coupler (Valenziano
et al. 2009). The signal is then amplified by a
factor
and shifted in phase by 0-180 degrees at 8 kHz
synchronously with the acquisition electronics. Finally, a second
hybrid coupler separates the input sky signal from the reference load
signal.
1.75 m long waveguides (D'Arcangelo et al. 2009a) are connected to the FEM in bundles of four elements providing the thermal break between 20 K and 300 K where the ambient temperature back-end section of the radiometer is located (Varis et al. 2009; Artal et al. 2009). The waveguides are thermally attached to the three thermal shields of the satellite, the V-grooves. They act as radiators to passively cool down the payload to about 50 K. They drive the thermal gradient along the waveguides.
Inside the BEM the signal is further amplified by
and then
detected by four output detector diodes
.
In nominal conditions, each of the four diodes detects a voltage
alternatively (each
s
which corresponds to 1/8192 Hz-1)
proportional by a factor a to the sky load
and reference load temperature. By differencing these two signals a
very stable output is obtained, which allows the measurement of very
faint signals.
Assuming negligible mismatches between the two radiometer legs
and within the phase switch, the differenced radiometer output at each
detector averaged over the bandwidth
can be written in terms of overall gain,
in units of
V/K, and noise temperature,
,
as
Here we further define the waveguide losses as



The ohmic losses of the feedhorn - OMT assembly,
,
and of the 4K Reference load system,
,
modify the actual sky and reference load through the following
equations
where

The r factor in Eq. (1) is the gain
modulation factor calculated as
which nulls the radiometer output. A more general form of the averaged power output, which takes into account various non-ideal behaviors of the radiometer components, can be found in Seiffert et al. (2002) and Mennella et al. (2003). The key parameters needed to reconstruct the required signal (differences of


2.2 Signal model
To better understand the purpose of the calibration, it is useful to write Eqs. (1) to (4) appropriately so that the attenuation coefficients are taken into account in the RCA parameters instead of considering their effects as a target effective temperature. For the sky signal the output can be written asand equivalently for the reference signal
Differencing Eqs. (5) and (8) we obtain the differenced (sky - ref) output similar to Eq. (1)
Equations (5), (8), and (11) are the basis for the RCA calibration, because all parameters involved were measured during the RCA test campaign by stimulating each RCA at cryo temperature (close to the operational in-flight conditions) with several known


2.3 Radiometer chain assembly calibration plan
Each RCA calibration included (i) functional tests, to verify the functionality of the RCA; (ii) bias tuning, to set the best amplifier gains and phase switch bias currents for maximum performance (i.e. minimum noise temperature and best radiometer balancing); (iii) basic radiometer property measurements to estimate G*,

Table 1: Calibration test list.
Although the calibration plan was the same for all RCAs, the differences in the setup between 30/44 GHz and 70 GHz resulted in a different test sequence and procedures, which retained the objective of RCA calibration unchanged.
At 70 GHz the RCA tests were carried out in a dedicated
cryogenic chamber developed by DA-Design
(formerly Ylinen Electronics), which was capable of accommodating two
RCAs at one time. Thus the RCA test campaign was planned for three RCA
pairs, namely RCA18 and RCA23,
RCA19 and RCA20, RCA21
and RCA22. At 30 and 44 GHz,
only one RCA at a time was calibrated. Due to the different length of
the waveguides, the thermal interfaces were not exactly the same for
different RCAs, which resulted in a slightly different thermal behavior
of the cryogenic chamber and thus slightly different calibration
conditions.
3 Radiometer chain assembly calibration facilities
In both cases, the calibration facilities used the cryogenic chamber
described in Terenzi
et al. (2009b), a calibration load, the skyload (Terenzi et al. 2009a),
electronic ground support equipment (EGSE) and software (Malaspina et al. 2009).
The heart of the EGSE was a breadboard of the LFI flight data
acquisition electronics (DAE) with Labwindow
software to control the power supplies to the FEMs and BEMs, read all
the housekeeping parameters and digitize the scientific signal at
8 KHz without any average or time integration.
The EGSE sent data continuously to a workstation operating the Rachel
(RAdiometer CHain EvaLuator) software for quick-look analysis and data
storage (Malaspina
et al. 2009). The data files were stored in FITS
format. As two chains were calibrated at the same time at
70 GHz, separate EGSEs and analysis workstations were used for
each RCA. Below the cryofacilities and skyloads are summarized with the
emphasis on the issues related to the analysis of the calibration data.
3.1 The cryofacility for the 30 and 44 GHz RCAs
The chamber with its overall dimensions of
![]() |
Figure 2: Radiometer chain assembly integrated into the 30 and 44 GHz cryofacility for calibration. In the picture at the top the skyload facing the horn is visible together with the FEM insulated from the 50 K shroud (the copper box). In the bottom picture the BEM and its thermal interface are shown. See the text for details of the cryochamber. |
Open with DEXTER |
During the RCA27 and RCA28
calibrations an uncertainty in the reference targets' temperature was
experienced. A visual inspection of the cryochamber after the RCA28
test gave a possible
explanation and in the RCA27 test, an additional
sensor was put on the back of one of the reference targets in order to
verify the probable source of the problem.
The observed behavior was consistent with an excess heat flow through
the 4 K reference load (4KRL), via its insulated support
caused by a contact created during cooldown. A dedicated thermal model
was thus developed to derive the Eccosorb 4KRL temperature,
from the
back plate controller sensor temperature,
(Terenzi et al. 2009b).
A quadratic fit was found with
for
each pair of detectors coupled to the same radiometer arm and for
each 30 GHz RCA. The coefficients derived from the fit are
shown in Table 2.
Table 2: Reference load target temperature.
3.2 Sky load at 30 and 44 GHz
The calibrator consisted of a cylindrical cavity with walls covered in Eccosorb CR110![[*]](/icons/foot_motif.png)




![]() |
Figure 3:
Thirty minutes of data acquired during |
Open with DEXTER |
Due to a failure in the sensor on the pyramids an analytical evaluation
of
temperature from
temperature was performed during the calibration of RCA24
and RCA27. The data are shown in Fig. 4. Although
the data show a linear behavior, the differences between
and
decrease as
the temperature increases (see Fig. 5) as expected
from the thermal behavior of the system, suggesting that a quadratic
fit with
is
more representative. This quadratic fit was performed, and the
coefficients are reported in Table 3.
Table 3: Temperarture of the sky load pyramids. Quadratic fit coefficients.
![]() |
Figure 4:
|
Open with DEXTER |
![]() |
Figure 5:
Differences between |
Open with DEXTER |
![]() |
Figure 6: Bologna design of the RCA sky load calibrator. The overall dimensions in mm are reported in the drawing on the left. The pyramids on the bottom of the skyload are clearly visible on the right picture. |
Open with DEXTER |
3.3 The cryofacility of the 70 GHz RCAs
This cryofacility has the dimensions
.
The facility has a layout similar to that at 30-44 GHz,
although the 70 GHz facility was designed to house two
radiometer chains simultaneously (Fig. 7).
The smaller dimensions of the feedhorns and front end modules and the decision to use two small dedicated sky loads directly in front of the horns allowed the cold part of the two RCAs under test to be contained in a volume similar to that of the 30 and 44 GHz chamber. Temperature interfaces such as FEMs, sky load and reference load were coupled together by means of copper slabs and then connected to the 4 K and 20 K coolers. The FEMs were controlled at their nominal temperature of 20 K; sky and reference loads were controlled in the range 10-25 K with a stability better than 10 mK; the back end modules were insulated from the chamber envelope by means of a supporting structure without temperature control, which was considered unnecessary.
![]() |
Figure 7: Top: two 70 GHz RCAs integrated in the Ylinen Electronics cryofacility. The two BEMs (on the right) are connected to the waveguides, here surrounded by aluminum mylar. On the left the shroud contains the two horns facing the ``Ylinen'' skyload at about 50 K. Bottom: detail of the front end. The two FEMs and the pair of horns are facing the two skyload containers. |
Open with DEXTER |
3.4 The skyloads at 70 GHz
The design for the 70 GHz RCA skyload was made by Ylinen Electronics. The basic design is shown in Fig. 8. The load configuration is a single folded conical structure in Eccosorb mounted in an aluminum housing. It is attached to a brass back plate. A single waveguide input is mounted through the back plate, providing the method of applying RF stimulus signals through the absorber for the RCA_SPR test (see Sect. 4.5). Load performances were measured over the whole V-band showing a return loss better than -20 dB. Two sensors were placed on the sky load, one at the controller stage, referred to as

![]() |
Figure 8: Ylinen design of the RCA sky load calibrator. This design produced a load with -20 dB of return loss over the whole bandwidth. The two light blue shaded regions represent the horn mouth at the left and the rectangular waveguide injector on the right. The absorber is enclosed in a metallic box except for the part facing the horn, which is closed with a Teflonplate. |
Open with DEXTER |
4 Methods and results
4.1 Functional tests
Functional tests were performed at ambient and at cryo temperature. All the RCAs were biased with nominal values and the power consumption was verified. In addition, each phase switch was operated in the nominal mode to check its functionality. As an example the functional test performed at cryogenic temperature on RCA26 is shown in Fig. 9. The figure reports the output voltage of the detector S-10 when the functional test is run: the BEM is switched on, the FEM is biased at nominal conditions and the fast 4 KHz switching is activated on the phase switches. It is evident that most of the change in signal is experienced when the BEM is on and the phase switches are biased correctly. These functional tests were also used as a reference for further tests up to the satellite-level verification campaign, besides checking the functionality of the RCA to proceed with the calibration.![]() |
Figure 9: Functional test performed at cryogenic temperature on RCA26. The two lines (orange and cyan) refer to the sky and load signals, when the 4 KHz switching is activated. Outside the interval 310-350 s the two curves are indistinguishable. |
Open with DEXTER |
4.2 Tuning
Before tuning the RCA, the DAE was set-up to read the voltage from each detector of the BEM,

![]() |
(14) |
The DAE gain,




The aim of the RCA tuning procedure (RCA_TUN)
was to choose the best bias conditions for each FEM low noise
amplifiers' (LNA) gate voltage and phase switch current. Each of the
four LNAs in a 30 GHz FEM consists of four amplification
stages (five for the FEM at 44 GHz), each driven by the same
drain voltage, .
The gate voltage
biases the first amplification stage, while
biases the
successive three (or four) stages. The phase switches are
driven by two currents (I1
and I2), biasing each diode.
The currents determine the amount of attenuation by each diode and thus
are adjusted to obtain the final overall radiometer balance.
The phase switches between the LNA and the second hybrid use
the interconnection of two hybrid rings to improve the bandwidth and
the matching with two Shunt PIN diodes. Depending on the polarization
of the diodes the signal travels into a circuit, which can be /2 longer,
so that it is shifted by
.
Details can be found in Hoyland
(2004) and in Cuttaia
et al. (2009). In Fig. 10 we report
the conceptual schematic diagram of the phase switch.
![]() |
Figure 10:
Conceptual scheme of the phase switch integrated into the radiometers.
Each phase switch is composed of two diodes commanded by the
currents I1 and I2. They act as a on/off switch.
Depending on the polarization state of the diodes the signal follows
the magenta path or the cyan |
Open with DEXTER |
At 30 and 44 GHz the phase switches were tuned with one radiometer leg
switched off. In these conditions the signal entering each phase switch
diode is the same, and the output signal at the DAE can be used
directly to precisely balance the two states of the phase switch. Any
differences in the sky and ref signal are caused only by the phase
switches and not by other non-idealities of the radiometers, nor by
different input target temperatures. The two currents were chosen to
minimize the quadratic differences,
,
between odd and even samples of the signal (corresponding in
Fig. 10
to the magenta and cyan path respectively). If for example the phase
switch was tuned on the same leg as the amplifier M1, the following
expression was minimized:
![]() |
(15) |
where S00, S01 are the two DAE outputs related to the ``M'' half FEM. In this case o and e refer to odd and even signal samples. The same differences for the other phase switches,



The front-end LNAs were tuned for noise temperature
performance, Tn,
and isolation, I. For each channel Tn
and I were measured as a function of the gate
voltages
and
.
Firstly the minimum noise conditions were found by varying
while keeping
and
fixed. The noise temperature was measured with the Y-factor
method (see Appendix A
for the details of this method). Because only relative estimates of Tn
are relevant for tuning purposes, we did not correct for the effect of
non-linearity in the 30 and 44 GHz RCAs.
Once the optimum
was determined, the optimum
was found by maximizing the isolation, I,
measured by varying




At 70 GHz the best working conditions were found by measuring
the
noise temperature and the isolation as a function of the gate voltages
and
,
but with a slightly different approach mainly for schedule reasons to
that of the lower frequency chains. The procedure required two
different temperatures for the
reference load (about 10 K in the low state and about
20 K in the high
state) and
,
,
and
were varied independently on the half FEM. Then the procedure was
repeated with both FEM legs switched on. The noise temperature was
measured with the Y-factor method from the signals
coming from the two temperature states, and the isolation was
calculated with Eq. (16).
Although the bias parameters found during the RCA tuning are the optimal ones, we found that different electrical and cryogenic conditions induce uncertainties in the bias values. This is due to the different grounding and the impact of the thermal gradient along the bias cables. To overcome this problem, tuning verification campaigns are planned at LFI integrated level and inflight. In both cases the RCA bias values have been assumed as reference values.
4.3 Basic properties
The basic properties of the radiometers, namely noise temperature, isolation, gain, and linearity were obtained in a single test.The RCA_LIS test was performed varying
(and
subsequently
)
in steps while keeping the
(and subsequently
)
constant. In Fig. 11
we give the temperature range spanned during the tests. Due to the
different thermal behavior of each RCA the range was not the same for
all the chains. The brightness temperature was calculated from
temperature sensors located in the external calibrators (both sky and
reference) and the output voltages of the four detectors were recorded.
The main uncertainty was in the determination of the actual brightness
temperature seen by the radiometer. At 30 and 44 GHz
the brightness temperature of the skyload was derived from the
thermometer located inside the pyramids, from where the main thermal
noise emission originated. For 70 GHz both the backplate
thermometer and the absorber thermometer were used to derive the
brightness temperature. However, the backplate and the absorber
temperatures were found to introduce a significant systematic error in
the reconstructed physical temperature of the load, as explained in
Sect. 3.4.
It was decided to calibrate the 70 GHz RCAs using the
reference load steps instead.
We denote here for simplicity the value of either
or
with
in
Eqs. (5)
and (8),
alternatively
or
with
,
and the corresponding
or
with
.
With G we denote the corresponding total
gain.
For a perfectly linear radiometer the output signal can be
written as
![]() |
(17) |
and the gain and system temperature can be calculated by measuring the output voltage for only two different values of the input temperature (Y-factor method). This was indeed the case for the 70 GHz RCAs. For the 30 and 44 GHz RCAs the determination of the basic properties was complicated by a significant non-linear component in the response of the 30 and 44 GHz RCAs. This was discovered during the previous qualification campaign and has been well characterized during these flight model RCA tests. The non-linearity effects in LFI are discussed in Mennella et al. (2009) together with its impact on the science performances. For a radiometer with compression effects the radiometer gain, G, is a function of total input temperature,

Particular care was required in the determination of the noise temperature of the 30 and 44 GHz RCAs. To overcome the problem, the application of four different types of fit were performed:
![]() |
Figure 11: Physical temperature ranges for the RCA_LIS tests. Black lines refer to the reference load temperature steps. Gray lines refer to the skyload temperature steps. |
Open with DEXTER |
![]() |
Figure 12: Typical temperature behavior of sky load (blue), reference load (green) and FEM body (red) during the RCA_STn test. The temperature stability is better than 1 mK. |
Open with DEXTER |
- 1.
- Linear fit. This fit was always
calculated as a reference, even for non-linear behavior of the
radiometer, so that
(19)
The linear gain,, and the noise temperature were derived. The fit was applied to all available data, not only to the two temperature steps as in the Y-factor method, to reduce the uncertainties for the linear 70 GHz RCAs and to evaluate the non-linearity of the 30 and 44 GHz chains.
- 2.
- Parabolic fit. This was applied to
understand the effect of the non-linearity, for the evaluation of which
the quadratic fit is the simplest way. The output of the fit were the
three coefficients from the equation
(20)
In this case the noise temperature was employed as the solution of the equation.
- 3.
- Inverse parabolic fit. This was used
because the noise temperature was directly derived from
(21)
where.
- 4.
- Gain model fit. A new gain model was
developed based on the results of Daywitt
(1989), modified for the LFI (see Appendix B). The
total power output voltage was written as
(22)
where G0 is the total gain in the case of a linear radiometer, b is the linear coefficient (b=0 in the linear case,for complete saturation, i.e. G(T) =0).

Table 4: Receiver basic properties.
Table 5: Receiver basic properties: gain, noise temperature, and linearity.
4.4 Noise properties
Radiometer noise properties were derived from the RCA_STn
test. This test consisted of acquiring data under stable thermal
conditions for at least three hours. Then the temperature of the loads
were changed to measure the noise properties at different sky and
reference target temperatures. As expected, the best 1/f
conditions were found when the difference between the sky and reference
load temperatures was minimal. This occurred at the first and last step
as seen in Fig. 12,
which represents a typical temperature profile of the test.
The amplitude spectral density was calculated for each output diode.
The 1/f component (knee frequency, ,
and slope,
),
the white noise plateau and the gain modulation factor, r,
were derived. At 70 GHz the 1/f spectrum
was clearly dominated by thermal instabilities of the BEM which was not
controlled in temperature. There the measured knee frequency values
were over-estimated, while at 30 and 44 GHz the
cryofacility was sufficiently stable to characterize the 1/f
performances of the radiometers. From the white noise and DC level the
effective bandwidth was calculated as
![]() |
(23) |
where





![]() |
Figure 13: Log-log plot of the amplitude spectral density of the differential detector output noise. The left plot refers to the RCA27M-00 detector with the sky and the reference loads both at 20 K; the plot in the center refers to the RCA26M-00 detector with sky and reference loads at 8 K and 13 K respectively; the plot on the right refers to the RCA23S-11 with the sky and reference loads at 15 K and 9 K respectively. |
Open with DEXTER |
Table 6: Receiver noise properties: 1/f knee frequency, slope, and r factor.
Apart from 1/f noise and white noise, spikes were observed in all RCAs: at 70 GHz they were caused by the electrical interaction between the two DAE units, which were slightly unsynchronised; at 30 and 44 GHz they were due to the housekeeping acquisition system. Because it was clear that the spikes were always due to the test setup and not to the radiometers themselves, the spikes were not considered critical at this stage, even if they showed up in frequency and amplitude.
As an example of the dependence of the noise performance on
the temperature the antenna temperature pairs used during the the tests
of the RCA28 are reported in Table 7. These
temperatures were calculated with the coefficients reported in
Tables 2
and 3
of Sect. 3.1
with the physical temperatures converted into antenna temperatures. The
differences between the
and the
were calculated for each arm of the radiometer. The resulting 1/f
knee frequency, the slope of the 1/f spectra,
and the gain modulation factor, r, are reported in
Fig. 14
as a function of the temperature differences. It is evident from these
plots that the knee frequency is increasing with the temperature
difference, as expected. Moreover, the gain modulation factor is
approaching unity as the input temperature difference becomes zero,
which agrees with Eq. (4).
The slope,
,
does not show any correlation with the temperature differences, because
it depends on the amplifiers
rather than on the
.
This behavior was also found in the other RCAs.
4.5 Bandpass
A dedicated end-to-end spectral response test, RCA_SPR, was designed and carried out to measure radiometer RF bandshape in operational conditions, i.e., on the integrated RCA with the front-end at the cryogenic temperature. An external RF source was used to inject a monochromatic signal sweeping through the band into the sky horn. Then the DC output of the radiometer was recorded as a function of the input frequency, giving the relative overall RCA gain-shape,
![]() |
(24) |
Different setup configurations were used. At 70 GHz the RF signal was directly injected into the sky horn. The input signal was varied by 50 points from 57.5 GHz to 82.5 GHz. At 30 GHz and 44 GHz the RF signal was injected into the sky horn after a reflection on the sky load absorber's pyramids, scanning in frequency from 26.5 GHz to 40 GHz in 271 points and from 33 GHz to 50 GHz in 341 points. The flexible waveguides WR28 and WR22 were used to reach the skyload for the 30 and 44 GHz RCAs (Fig. 15). The input signal was not calibrated in power because only a relative band shape measurement was required. The stability of the signal was ensured by the use of a synthesized sweeper generator guaranteeing the stability of the output within 10%. The attenuation curve of the waveguide carrying the signal from the sweeper to the injector was treated as a rectangular standard waveguide with losses during the data analysis.
![]() |
Figure 14:
1/f knee frequency (asterisk on the left), gain
modulation factor (diamonds in the center), and
the slope of the 1/f spectrum (triangles on the
right) as a function of
|
Open with DEXTER |
![]() |
Figure 15: Setup of the RCA_SPR tests. The picture and the sketch on the left report the test setup of the 30 and 44 GHz RCAs. The flexible waveguide is clearly visible on the picture on the side of the horn. The picture and the sketch on the right report the setup of the 70 GHz RCAs. There the signal was injected infront of the horn through the sky load, and copper rigid waveguides were used to carry the signal form the generator to the RCAs. |
Open with DEXTER |
All RCA bandshapes were measured, but for the two 30 GHz RCAs
only half a radiometer was successfully tested due to a setup problem
that appeared when the RCAs were cooled down. For schedule reasons it
was not possible to repeat the test at the operational temperature, and
only a check at the warm temperature was performed. This warm test was
not used for calibration due to different dynamic range, amplifier
behavior, and bias conditions. Results are reported in Table 8 and plots of all
the measurements in Figs. 16-18. All curves
reported in the plots are normalized to the area so that
![]() |
(25) |
The bandshape is mainly determined by the filter located inside each BEM, whose frequency response is independent of the tuning of the FEM amplifiers. The dependence of the bandshape on the amplifier biases has been checked on the 30 GHz radiometers (De Nardo 2008), showing that at first order the response remains unchanged. A similar situation occurs on the RCAs at 44 GHz and 70 GHz, where the tuning has second order effects on the overall frequency response.
![]() |
Figure 16: Measured relative gain function (bandpass) of the 8 detectors at 30 GHz. The curves that show big ripples are those caused by the setup problem (see text). The bandpasses are normalized to the area as explained in the text. |
Open with DEXTER |
![]() |
Figure 17: Measured relative gain function (bandpass) of all 12 detectors at 44 GHz. The bandpasses are normalized to the area as explained in the text. |
Open with DEXTER |
![]() |
Figure 18: Measured relative gain function (bandpass) of all 24 detectors at 77 GHz. The bandpasses are normalized to the area as explained in the text. |
Open with DEXTER |
4.6 Susceptibility
Any variation in physical temperature of the RCA,

where


Table 7: Receiver noise properties. Long duration test antenna temperature pairs for the RCA28.
Table 8: Spectral response test results.
Table 9: The transfer function of the susceptibility of FEM to temperature variations.
The susceptibility of the radiometer signal to temperature
variations in the BEM and 3
V-groove were measured only for the 30 and 44 GHz
chains, because at 70 GHz it was not possible to control the
temperatures of these interfaces in their cryofacility. Here we report
on the BEM susceptivity tests only: it is the prominent radiometric
effect between both, because the diodes are thermally attached to the
BEM body. The total power output voltage on each BEM detector can then
be expressed by modifying Eq. (18)
![]() |
(27) |
There




![]() |
(28) |
with the transfer function,

![]() |
(29) |
The values are reported in Table 10.
Table 10: The transfer function of the susceptibility of BEM to temperature variations.
5 Conclusions
The eleven LFI RCAs were calibrated according to the overall LFI calibration plan. The front end low-noise amplifiers and phase switches were properly tuned to guarantee minimum noise temperature and best isolation. Basic performances (noise temperature, isolation, linearity, photometric gain), noise properties (1/f spectrum, noise equivalent bandwidth), relative bandshape, and susceptibility to thermal variations were measured in a dedicated cryogenic environment as close as possible to flight-operational conditions. All radiometric parameters were measured with excellent repeatability and reliability, except for 1/f noise at 70 GHz and some of the bandpasses at 30 GHz. The measurements were essentially in line with the design expectations, indicating a satisfactory instrument performance. Ultimately all RCA units were accepted, because the measured performances were in line with the scientific goal of LFI. The RCA test campaign described here represents the primary calibration test for some key radiometric parameters of the LFI, because they were not accurately measured as part of the RAA calibration campaign, nor can they be measured in-flight during the Planck nominal survey:- Tuning results were used to set the subsequent tuning verification procedure up to the calibration performance and verification phase (CPV) in flight.
- Non-linear behavior of the 30 and 44 GHz RCAs was accurately measured and characterized and used to estimate the impact in flight (Mennella et al. 2009). Moreover each radiometer system noise temperature was accurately determined.
- Except for some of the measured spectral responses, in which systematic effects arising from standing waves in the sky load were experienced, the RCA band shapes were only measured and characterized during the RCA tests. Together with the independent estimates of the band shape based on a synthesis of measured responses at unit level (Zonca et al. 2009), they are essential for the flight data analysis.
- Susceptivity to thermal variation of the FEM and BEM was measured and represents the reference values because only the FEM susceptivity was measured during RAA tests (Terenzi et al. 2009a), but under worse thermal conditions.
Planck is a project of the European Space Agency with instruments funded by ESA member states, and with special contributions from Denmark and NASA (USA). The Planck-LFI project is developed by an International Consortium lead by Italy and involving Canada, Finland, Germany, Norway, Spain, Switzerland, UK, USA. The Italian contribution to Planck is supported by the Italian Space Agency (ASI). In Finland, the Planck LFI 70 GHz work was supported by the Finnish Funding Agency for Technology and Innovation (Tekes) T.P.'s work was supported in part by the Academy of Finland grants 205800, 214598, 121703, and 121962. T.P. thank Waldemar von Frenckells stiftelse, Magnus Ehrnrooth Foundation, and Väisälä Foundation for financial support. The Spanish participation is funded by Ministerio de Ciencia e Innovacion through the projects ESP2004-07067-C03-01 and AYA2007-68058-C03-02.
Appendix A: Y-factor method
The classical Y-factor method is the simplest way
to calculate the noise temperature, and it requires that radiometric
data are acquired at two different (possibly well-separated) physical
temperatures of one of the input loads. Below we will assume a sky load
temperature increase. Clearly the treatment is completely symmetrical
if the reference load temperature is increased.
If we denote with
and
as the antenna temperatures corresponding to the skyload physical
temperatures, we find that the ratio between the output voltages is
given by
![]() |
(A.1) |
The system noise temperature is then calculated as
![]() |
(A.2) |
Appendix B: Radiometer non-linear model (gain model)
A non-linear gain model was developed and applied to the 30 and 44 GHz RCAs to model the observed behavior of the output voltages as a function of input temperature. The model was developed on the basis of Daywitt (1989) and specifically adopted for the LFI 30 and 44 GHz RCAs, which exhibit a non-negligible compression effect in the BEMs. The FEM is assumed to have a constant gain and noise temperature![]() |
(B.1) |
The BEM (Artal et al. 2009) shows an overall gain (including the detector diode), which depends on the BEM input power as follows:
![]() |
(B.2) |
where p is the power entering the BEM and b is a parameter defining the non-linearity of the BEM. For b=0 the radiometer is linear; for



The power entering the BEM (we here neglect the attenuation of
the waveguides whose effect can be modeled as a small reduction of the
FEM gain and a small increase of the FEM noise temperature) can be
written as
![]() |
(B.3) |
where
![]() |
(B.4) |
The voltage at each output BEM detector (the detector diode constant is included in the BEM gain) can be written as
In a compact way Eq. (B.6)
can be written as
![]() |
(B.7) | ||
![]() |
(B.8) |
where the

References
- Artal, E., Aja, B., de la Fuente, M. L., et al. 2009, JINST, 4, T12003 [NASA ADS] [Google Scholar]
- Bersanelli, M., Mandolesi, N., Butler, R. C., et al. 2010, A&A, 520, A4 [NASA ADS] [CrossRef] [EDP Sciences] [Google Scholar]
- Cuttaia, F. 2005, Ph.D. Thesis, University of Bologna [Google Scholar]
- Cuttaia, F., Mennella, A., Stringhetti, L., et al. 2009, JINST, 4, T12013 [NASA ADS] [Google Scholar]
- D'Arcangelo, O., Figini, L., Simonetto, A., et al. 2009a, JINST, 4, T12007 [NASA ADS] [Google Scholar]
- D'Arcangelo, O., Simonetto, A., Figini, L., Pagana, E., Villa, F., Pecora, M., Battaglia, P., Bersanelli, M., Butler, R. C., Garavaglia, S., Guzzi, P., Mandolesi, N., & Sozzi, C. 2009b, JINST, 4, T12005 [Google Scholar]
- Davis, R. J., Wilkinson, A., Davies, R. D., et al. 2009, JINST, 4, T12002 [NASA ADS] [Google Scholar]
- Daywitt, W. C. 1989, NIST Tech. Note, 1327 [Google Scholar]
- De Nardo, S. 2008, Master's Thesis, Univeristà degli Studi di Milano [Google Scholar]
- Hoyland, R. 2004, US patent 6, 803, 838 B2 [Google Scholar]
- Malaspina, M., Franceschi, E., Battaglia, P., et al. 2009, JINST, 4, T12017 [NASA ADS] [Google Scholar]
- Mennella, A., Bersanelli, M., Seiffert, M., et al. 2003, A&A, 410, 1089 [NASA ADS] [CrossRef] [EDP Sciences] [Google Scholar]
- Mennella, A., Villa, F., Terenzi, L., et al. 2009, JINST, 4, T12011 [NASA ADS] [Google Scholar]
- Mennella, A., Bersanelli, M., Butler, R. C., et al. 2010, A&A, 520, A5 [NASA ADS] [CrossRef] [EDP Sciences] [Google Scholar]
- Seiffert, M., Mennella, A., Burigana, C., et al. 2002, A&A, 391, 1185 [NASA ADS] [CrossRef] [EDP Sciences] [Google Scholar]
- Terenzi, L., Cuttaia, F., De Rosa, A. L. V., et al. 2009a, JINST, submitted [Google Scholar]
- Terenzi, L., Lapolla, M., Laaninen, M., et al. 2009b, JINST, 4, T12015 [NASA ADS] [Google Scholar]
- Terenzi, L., Salmon, M. J., Colin, A., et al. 2009c, JINST, 4, T12012 [Google Scholar]
- Valenziano, L., Cuttaia, F., De Rosa, A., et al. 2009, JINST, 4, T12006 [Google Scholar]
- Varis, J., Hughes, N. J., Laaninen, M., et al. 2009, JINST, 4, T12001 [NASA ADS] [Google Scholar]
- Villa, F., D'Arcangelo, O., Pecora, M., et al. 2009, JINST, 4, T12004 [NASA ADS] [Google Scholar]
- Zonca, A., Franceschet, C., Battaglia, P., et al. 2009, JINST, 4, T12010 [Google Scholar]
Footnotes
- ...
- Present address: Astrium GmbH, Friedrichshafen, Germany.
- ... mission
- Planck (http://www.esa.int/Planck) is a project of the European Space Agency - ESA - with instruments provided by two scientific Consortia funded by ESA member states (in particular the lead countries: France and Italy) with contributions from NASA (USA), and telescope reflectors provided in a collaboration between ESA and a scientific Consortium led and funded by Denmark.
- ... 70 GHz
- The chains are numbered as RCAXX where XX is a number from 18 to 23 for the 70 GHz RCAs, from 24 to 26 for the RCAs at 44 GHz, and from 27 to 28 for the RCAs at 30 GHz.
- ... diodes
- According to the name convention, diodes refer to the Main-polarization are labeled as M-00, M-01 and those referring to the Side-polarization are labeled as S-10, S-11.
- ... DA-Design
- http://www.da-design.fi/space
- ...
- http://www.ni.com/lwcvi/
- ... CR110
- Emerson & Cuming, http:www.eccosorb.com
All Tables
Table 1: Calibration test list.
Table 2: Reference load target temperature.
Table 3: Temperarture of the sky load pyramids. Quadratic fit coefficients.
Table 4: Receiver basic properties.
Table 5: Receiver basic properties: gain, noise temperature, and linearity.
Table 6: Receiver noise properties: 1/f knee frequency, slope, and r factor.
Table 7: Receiver noise properties. Long duration test antenna temperature pairs for the RCA28.
Table 8: Spectral response test results.
Table 9: The transfer function of the susceptibility of FEM to temperature variations.
Table 10: The transfer function of the susceptibility of BEM to temperature variations.
All Figures
![]() |
Figure 1: Scheme of the RCA and its flight-like thermal interfaces. The FEM is at 20 K, while the BEM is at 315 K. At 30 and 44 GHz the thermal interface attached to the third and coldest V-groove (VG-3) is controlled at a temperature near 60 K, while for the 70 GHz RCA the VG-3 interface was not controlled in temperature. Both loads - the sky load and the reference load - are controlled in a temperature of approximately 4 K to 35 K. |
Open with DEXTER | |
In the text |
![]() |
Figure 2: Radiometer chain assembly integrated into the 30 and 44 GHz cryofacility for calibration. In the picture at the top the skyload facing the horn is visible together with the FEM insulated from the 50 K shroud (the copper box). In the bottom picture the BEM and its thermal interface are shown. See the text for details of the cryochamber. |
Open with DEXTER | |
In the text |
![]() |
Figure 3:
Thirty minutes of data acquired during |
Open with DEXTER | |
In the text |
![]() |
Figure 4:
|
Open with DEXTER | |
In the text |
![]() |
Figure 5:
Differences between |
Open with DEXTER | |
In the text |
![]() |
Figure 6: Bologna design of the RCA sky load calibrator. The overall dimensions in mm are reported in the drawing on the left. The pyramids on the bottom of the skyload are clearly visible on the right picture. |
Open with DEXTER | |
In the text |
![]() |
Figure 7: Top: two 70 GHz RCAs integrated in the Ylinen Electronics cryofacility. The two BEMs (on the right) are connected to the waveguides, here surrounded by aluminum mylar. On the left the shroud contains the two horns facing the ``Ylinen'' skyload at about 50 K. Bottom: detail of the front end. The two FEMs and the pair of horns are facing the two skyload containers. |
Open with DEXTER | |
In the text |
![]() |
Figure 8: Ylinen design of the RCA sky load calibrator. This design produced a load with -20 dB of return loss over the whole bandwidth. The two light blue shaded regions represent the horn mouth at the left and the rectangular waveguide injector on the right. The absorber is enclosed in a metallic box except for the part facing the horn, which is closed with a Teflonplate. |
Open with DEXTER | |
In the text |
![]() |
Figure 9: Functional test performed at cryogenic temperature on RCA26. The two lines (orange and cyan) refer to the sky and load signals, when the 4 KHz switching is activated. Outside the interval 310-350 s the two curves are indistinguishable. |
Open with DEXTER | |
In the text |
![]() |
Figure 10:
Conceptual scheme of the phase switch integrated into the radiometers.
Each phase switch is composed of two diodes commanded by the
currents I1 and I2. They act as a on/off switch.
Depending on the polarization state of the diodes the signal follows
the magenta path or the cyan |
Open with DEXTER | |
In the text |
![]() |
Figure 11: Physical temperature ranges for the RCA_LIS tests. Black lines refer to the reference load temperature steps. Gray lines refer to the skyload temperature steps. |
Open with DEXTER | |
In the text |
![]() |
Figure 12: Typical temperature behavior of sky load (blue), reference load (green) and FEM body (red) during the RCA_STn test. The temperature stability is better than 1 mK. |
Open with DEXTER | |
In the text |
![]() |
Figure 13: Log-log plot of the amplitude spectral density of the differential detector output noise. The left plot refers to the RCA27M-00 detector with the sky and the reference loads both at 20 K; the plot in the center refers to the RCA26M-00 detector with sky and reference loads at 8 K and 13 K respectively; the plot on the right refers to the RCA23S-11 with the sky and reference loads at 15 K and 9 K respectively. |
Open with DEXTER | |
In the text |
![]() |
Figure 14:
1/f knee frequency (asterisk on the left), gain
modulation factor (diamonds in the center), and
the slope of the 1/f spectrum (triangles on the
right) as a function of
|
Open with DEXTER | |
In the text |
![]() |
Figure 15: Setup of the RCA_SPR tests. The picture and the sketch on the left report the test setup of the 30 and 44 GHz RCAs. The flexible waveguide is clearly visible on the picture on the side of the horn. The picture and the sketch on the right report the setup of the 70 GHz RCAs. There the signal was injected infront of the horn through the sky load, and copper rigid waveguides were used to carry the signal form the generator to the RCAs. |
Open with DEXTER | |
In the text |
![]() |
Figure 16: Measured relative gain function (bandpass) of the 8 detectors at 30 GHz. The curves that show big ripples are those caused by the setup problem (see text). The bandpasses are normalized to the area as explained in the text. |
Open with DEXTER | |
In the text |
![]() |
Figure 17: Measured relative gain function (bandpass) of all 12 detectors at 44 GHz. The bandpasses are normalized to the area as explained in the text. |
Open with DEXTER | |
In the text |
![]() |
Figure 18: Measured relative gain function (bandpass) of all 24 detectors at 77 GHz. The bandpasses are normalized to the area as explained in the text. |
Open with DEXTER | |