Table 1.

MAG performance requirements, including measurement requirements and constraints.

Requirement Measurement Constraint
Resolve fluid-scale structures – Cadence up to proton gyroperiod – Cadence of better than 0.1 s
throughout the mission – Precision of better than 10% of – Absolute precision of < 10 pT
ambient field
– Sensitivity lower than natural variability – Instrument noise density lower
than 10 pT/ at 1 Hz
– Continuous measurement – Instrument operational throughout
science phase, not saturated by ambient
field

Resolve ion kinetic-scale phenomena – Cadence > 10 times proton gyroperiod – Cadence of better than 0.01 s
throughout the mission in burst mode in burst mode
– Instrument in burst mode for around
10% of operations

Measure ambient field vector – Absolute value within 10% of ambient – Instrument offset stability < 0.5 nT
for other instruments per 100 h