Free Access
Table 1.
MAG performance requirements, including measurement requirements and constraints.
Requirement | Measurement | Constraint |
---|---|---|
Resolve fluid-scale structures | – Cadence up to proton gyroperiod | – Cadence of better than 0.1 s |
throughout the mission | – Precision of better than 10% of | – Absolute precision of < 10 pT |
ambient field | ||
– Sensitivity lower than natural variability | – Instrument noise density lower | |
than 10 pT/![]() |
||
– Continuous measurement | – Instrument operational throughout | |
science phase, not saturated by ambient | ||
field | ||
Resolve ion kinetic-scale phenomena | – Cadence > 10 times proton gyroperiod | – Cadence of better than 0.01 s |
throughout the mission | in burst mode | in burst mode |
– Instrument in burst mode for around | ||
10% of operations | ||
Measure ambient field vector | – Absolute value within 10% of ambient | – Instrument offset stability < 0.5 nT |
for other instruments | per 100 h |
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.