Table 2
HERMES main parameters.
Spectral resolution | |
HRF | 85 000 (with slicer) |
LRF | 63 000 (interlaced with simultaneous |
Thorium spectrum) | |
Fibre aperture | |
HRF | 2.5 arcsec (80-μm fibre) |
LRF | 2.15 arcsec (60-μm fibre) |
Sampling | |
HRF | 2–2.7 pixels |
LRF | 3.1–3.8 pixels |
Spectral range | 377–900 nm (in 55 orders) |
Spectrograph layout | White-pupil layout on 1.2 × 2.4 m bench |
Beam diameter | 152 mm |
Collimator focal length | 1400 mm (f/9.2) |
Echelle grating | θ = 69.74°, 52.676 gr/mm, 154 × 408 mm |
Cross disperser | 2 PBL1Y prisms, 37.4° apex angle |
Camera focal length | 475 mm (f/3.1, 6 lenses) |
Detector | 2048 × 4608 13.5-μm pixels, thinned BI |
CCD with graded AR coating | |
Peak efficiency | |
HRF | B: 21%, V: 28%, R: 25%, I: 17% |
LRF | B: 15%, V: 20%, R: 18%, I: 12% |
Limiting magnitudes | 1 h exposure |
S/NV = 20 | mV = 13.4 (HRF), mV = 13.2 (LRF) |
S/NV = 100 | mV = 10.4 (HRF), mV = 10.0 (LRF) |
Radial velocity error | ~2.5 m s-1 (short term) |