Up: Comparison of source detection images
8 Test 5 - completeness and confusion
In this section we investigate the confusion and completeness problems
for XMM-Newton shallow and deep observations like the first XMM Deep X-ray
Survey in the Lockman Hole (Hasinger et al. 2001).
A set of 10 images with exposure times of 10 ks and 100 ks in the
energy bands [0.5-2] and [2-10] keV were generated; the fluxes were
drawn using the latest
relations from Hasinger et al. (2001) and Giacconi et al. (2000). Detection and
analysis were performed with exactly the same parameters for all
simulations: detection threshold, analysis threshold, background map
size, detection likelihood, etc. (see Sect. 3).
Cross-identification was achieved using the input sources above 10
counts and 30 counts for 10 ks and 100 ks exposures respectively.
Lowering the count limits yields more cross-identifications but
increases considerably the number of spurious detections.
The input image for 100 ks and [0.5-2] keV band is shown in
Fig. 12. The inner
zone where all analysis
is performed is indicated, as well as the total XMM-Newton field-of-view.
It is informative to compare it with the images for 10 ks in
Fig. 8.
![\begin{figure}
\par\includegraphics[width=7cm,clip]{MS10417f12.eps} \end{figure}](/articles/aa/full/2001/17/aa10417/Timg63.gif) |
Figure 12:
Simulated 100 ks XMM-Newton deep field in [0.5-2] keV with the same
parameters (
,
background) as in
the Lockman Hole (Hasinger et al. 2001; Watson et al.
2001). We restrict the analysis to the inner
 |
In order to estimate the effect of confusion we have generated images
with only point-like sources, distributed on a grid such to avoid
close neighbours, and with fixed fluxes spanning the interval
[10-16,10-13] erg/s/cm2.
In the following we discuss some important points.
- Confusion and completeness
The detection rate of the input sources as a function of flux
(Fig. 13) indicates that confusion problems are
significant for 100 ks exposures in the [0.5-2] keV band.
They are
marginal for 100 ks exposures in [2-10] keV band and absent for 10 ks in both bands.
![\begin{figure}
\par\mbox{\includegraphics[width=8cm,clip]{MS10417f13a.eps} \incl...
...S10417f13c.eps} \includegraphics[width=8cm,clip]{MS10417f13d.eps} }
\end{figure}](/articles/aa/full/2001/17/aa10417/Timg64.gif) |
Figure 13:
Ratio of number of cross-identified objects to the number of
input objects, with counts greater than 10 and 30 for 10 ks and
100 ks exposures respectively, as a function of the input flux.
The results of 10 generations are shown by crosses, the heavy
histogram is the corresponding average; the curve indicates the
detection rate if confusion is absent and the dotted line marks
90% completeness |
Energy conversion factors and the limiting fluxes below which more
than 10% of the input objects are lost are shown in
Table 8;
Table 8:
The energy conversion factors (ECF) and the 90% completeness limits
for detections. ECF is computed assuming power-low spectrum with
photon index
1019 cm-2 (Hasinger et al. 2001) and the
three EPIC instruments (pn+2MOS) with thin filters
|
[0.5-2] keV |
[2-10] keV |
ECF (cts/s per erg/s/cm2) |
|
6.70 10-13 |
3.66 10-12 |
Flux limits (erg/s/cm2) |
10 ks |
2 10-15 |
10-14 |
100 ks |
6 10-16 |
3 10-15 |
- Differential flux distribution
The differential flux distributions for 100 ks in [0.5-2] keV
obtained by MR/1+SE and EMLDETECT are shown in Fig. 14.
Spurious detections appears to be numerous with EMLDETECT below
5 10-16 erg/s/cm2 and tend to compensate the loss of sensitivity and
confusion. MR/1+SE appears to be less affected and allows us to set
a conservative flux limit of
6 10-16 erg/s/cm2 (
90 photons for 100 ks), below which the incompleteness becomes
important - 65% of the input sources are lost between
3 10-16 and
6 10-16 erg/s/cm2;
![\begin{figure}
\par\mbox{ \includegraphics[width=7cm,clip]{MS10417f14a.eps} \includegraphics[width=7cm,clip]{MS10417f14b.eps} }
\end{figure}](/articles/aa/full/2001/17/aa10417/Timg65.gif) |
Figure 14:
Differential number counts as a function of flux. The continuous
histogram is the input distribution, the dots with error bars
(Poissonian) are the measured distribution (without any
cross-identifications), the dashed histogram indicates the
false detections
and the continuous line is the
corrected distribution excluding
the false detections.
Left
for MR/1+SE, right for EMLDETECT
- all averaged from 10 simulations |
-
The
functions in [0.5-2] keV for 10 ks and 100 ks
exposures are shown in Fig. 15. The inferred
(by simple source counting) are in very good agreement
with the input ones down to fluxes about
2 10-15 and
6 10-16 erg/s/cm2 with MR/1+SE and 10-15 and
3 10-16 erg/s/cm2 for EMLDETECT for 10 ks and 100 ks respectively.
Although there are confusion and completeness problems for 100 ks
starting at
5-6 10-16 erg/s/cm2 as discussed above, their
effect is completely masked in the EMLDETECT integral distribution,
which seems to be in very good agreement down to
3 10-16 erg/s/cm2 (the lower flux limit for the Lockman Hole Deep Survey
analysis of Hasinger et al. 2001);
![\begin{figure}
\par\mbox{\includegraphics[width=7cm,clip]{MS10417f15a.eps} \includegraphics[width=7cm,clip]{MS10417f15b.eps} }
\end{figure}](/articles/aa/full/2001/17/aa10417/Timg66.gif) |
Figure 15:
The integral number of objects in the inner
(surface
0.087 sq.deg.) as a function of the flux for [0.5-2] keV and two
exposures: 10 ks and 100 ks. The points with error bars
(Poissonian) are the detections (without cross-identification)
with MR/1+SE boxes are EMLDETECT results, while the histogram is the
input
function |
- Photometric accuracy
The photometric reconstruction is relatively similar for 10 ks and
100 ks in for fluxes greater than
2 10-15 and
6 10-16 erg/s/cm2
respectively (the 90% completeness limit,
Fig. 16). The flux uncertainties are about 25-30%
and going down to fainter fluxes leads to very poor photometry.
![\begin{figure}
\par\mbox{\includegraphics[width=7cm,clip]{MS10417f16a.eps} \includegraphics[width=7cm,clip]{MS10417f16b.eps} }
\end{figure}](/articles/aa/full/2001/17/aa10417/Timg67.gif) |
Figure 16:
Photometry reconstruction for all 10 simulated images at 10 ks
(left) and 100 ks (right) in the [0.5-2] keV band. The solid
line is exact match between detected and input counts while the
dashed lines are for two-fold differences. The vertical dashed
line marks the 90% completeness limit (see
Table 8) and mean and st.dev. (in brackets) above
this limit are denoted |
Up: Comparison of source detection images
Copyright ESO 2001