A&A 393, 183-193 (2002)
DOI: 10.1051/0004-6361:20021020
P. Bordé1 - V. Coudé du Foresto1 - G. Chagnon1 - G. Perrin1
LESIA, FRE2461, Observatoire de Paris, 5 place Jules Janssen,
92195 Meudon, France
Received 12 March 2002 / Accepted 27 June 2002
Abstract
Long baseline stellar interferometry shares with other techniques
the need for calibrator stars in order to correct for instrumental and
atmospheric effects. We present a catalogue of 374 stars carefully selected
to be used for that purpose in the near infrared. Owing to several
convergent criteria with the work of Cohen et al. (1999), this
catalogue is in essence a subset of their self-consistent all-sky network of
spectro-photometric calibrator stars. For every star, we provide the angular
limb-darkened diameter, uniform disc angular diameters in the J, H and K bands, the Johnson photometry and other useful parameters. Most stars are
type III giants with spectral types K or M0, magnitudes V=3-7 and K=0-3.
Their angular limb-darkened diameters range from 1 to 3 mas with a median
uncertainty as low as 1.2%. The median distance from a given point on the
sky to the closest reference is
,
whereas this distance never
exceeds
for any celestial location.
Key words: catalogs - stars: fundamental parameters - instrumentation: interferometers - techniques: interferometric
A stellar interferometer collects and recombines light from separate
apertures. The interferometric observable, called the visibility, is a
measure of the spatio-temporal coherence of the target. It is a complex
quantity whose modulus and phase are respectively derived from the contrast
and the position of the fringe packet resulting from the recombination
process. In order to take into account instrumental and atmospheric
effects, observations of the target should be bracketed by observations of
calibrator stars. Thus, the object visibility
can be deduced
from the measured object and reference fringe contrasts
and
by
To improve calibration,
and
should
be measured with the same instrumental and preferably atmospheric
conditions, hence the need for a catalogue of calibrator stars with
a sky coverage as complete as possible. Therefore, in the framework of the
simplest single-parameter UD model, a dense grid of reference stars along
with their angular diameters has to be worked out. Since only few stellar
angular diameters have been directly measured so far at a precision
better than 5%, it seems unavoidable to rely also on indirect measurements.
Two approaches are possible: collecting all existing observations and
compiling them (e.g. the CHARM catalogue by Richichi & Percheron
2002), or selecting carefully a few hundred stars with well known
characteristics and deduced diameters. The first approach leads to a larger
but more heterogeneous catalogue with one entry per observation, whereas in
the second one, one star corresponds to one entry and all diameters are
given with comparable precision. The last option was chosen for the work
presented in this paper. It originated in the need of a grid of
calibrator stars among the FLUOR interferometry team. As
this need arises for every interferometer, we thought of making this grid
available to other interferometry teams through publication. FLUOR, the
Fiber Linked Unit for Optical Recombination, is currently hosted at the IOTA,
the Infrared and Optical Telescopes Array (Traub 2000), and is
described by Coudé du Foresto et al. (1998). The IOTA
features baselines between 7 and 38 m and FLUOR operates in the K and L bands.
In the next section, we describe the requirements that any star would
ideally fulfill to be selected as a calibrator. In Sect. 3, we briefly
review the work done by Cohen et al. (1999) (tenth paper of a
series, hereafter referred to as C.X) and we detail the way we extracted a
sublist of calibrator stars. Section 4 deals with statistical
characteristics of the stellar sample in terms of sky coverage or magnitude
and angular diameter distributions. Section 5 is dedicated to the
catalogue format, the calculation of the UD diameters and the definition
of some specific quantities. Finally, we compare angular diameter
determinations by C.X with existing measurements in Sect. 6.
We have argued that any source with a predictable visibility can be chosen as a reference. In that sense, a double star can be a good candidate provided all orbital parameters are known with enough accuracy. However for reliability as well as practical reasons, we prefer to reject everything but single non-variable stars with compact atmospheres. Indeed, variable stars are either multiple or change diameters with time, and extended atmospheres would lead to non trivial visibility models. In those cases, one may risk that the formal error bars would not represent the true errors because of the inadequacy of the chosen model. All this can be expressed in terms of the following observational constraints that remain to be quantitatively precised:
For their all-sky radiometric network, C.X have defined a set of criteria very close to those of Sect. 2. Moreover, they have derived angular diameters for all their stars, and found an excellent agreement between their predicted diameters and those measured by interferometry. It seems very attractive to use their network as the core of a catalogue of calibrators meant for interferometry. With this goal in mind, let us review briefly some key points of their work.
C.X have assembled composite spectra for a dozen giant stars from observed fragments in various spectral bands. These spectra are absolutely calibrated with respect to their primary IR standard, Sirius, and thus constitute a set of secondary standards. Fitting Kurucz model atmosphere (Kurucz 1993) to these spectra yields angular diameters with formal uncertainties. Now, C.X make the fundamental hypothesis that the deredenned spectrum of a giant star with a given spectral type (within the range K0-M0) can serve as a template for every other star with the same spectral type (the so-called "template assumption''). Then, infrared photometric measurements in several bands provide the correct scaling factor for the corresponding template. With the goal of finding at least one candidate per 50 deg2 on the sky, they have searched the IRAS database with the following criteria:
In addition to previously mentioned spectro-photometric criteria, the
interferometric context requires that a single star would be fed into the
collecting apertures. Therefore, acceptable stars could be part of a
multiple system provided the separations between the components are wider
than the interferometric field of view, or provided the interferometer would
not be able to resolve the system. For reliability reasons, we shall not
allow binary systems of the last category in this work. Since the
interferometric field of view of single-mode interferometers, like FLUOR, is
of the order of an Airy disc, angular separations of a few
could
theoretically be suitable. However, the pointing accuracy of the telescope,
the ability of the tracking system to remain locked on the right component
and the confusion risk between close stars should also be taken into
account. Therefore, considering a pointing accuracy of
and average seeing conditions (1-
), we have chosen to
discard all double stars with separations less than
,
regardless
of the difference in magnitude between the components. Concerning double
stars with both components in the field of view of the tracking system,
in the case of IOTA, we shall require a difference in visual
magnitude of 5 between the components (i.e. the secondary
contribute to 1% or less to the total visual flux).
We have investigated the nature of all the sources in Cohen's list with
the Simbad database and found that all stars belong to one of the twelve
types ordered in Table 1. When a new source which could be
classified under several object types is added to the Simbad database, a
unique choice is made according to the hierarchical classification proposed
by Ochsenbein & Dubois (1992). Table 1 states
the priority order for our twelve object types. For instance, a variable
star in a double system would be classified as variable, and a spectroscopic
binary in a cluster would be classified as a spectroscopic binary.
| Priority | Numeric | Standard name | Nb. of records |
| (Simbad) | (Total = 422) | ||
| 1 | 14.09.05.4 | Semi-regular pulsating star | 5 |
| 2 | 14.09.04.5 | Variable of RS CVn type | 3 |
| 3 | 12.13.01.2 | Eclipsing binary of Beta Lyrae type | 1 |
| 4 | 14.09.05.0 | Pulsating variable star | 2 |
| 5 | 14.09.01.0 | Variable of irregular type | 3 |
| 6 | 12.13.02.0 | Spectroscopic binaries | 12 |
| 7 | 14.09.00.0 | Variable star | 99 |
| 8 | 14.07.00.0 | High proper-motion star | 21 |
| 9 | 14.04.00.0 | Star in double system | 56 |
| 10 | 14.01.00.0 | Star in a cluster | 3 |
| 11 | 12.13.00.0 | Double or multiple star | 12 |
| 12 | 14.00.00.0 | Star | 205 |
Stars belonging to categories 1 to 5 (14.09.05.4, 14.09.04.5, 12.13.01.2,
14.09.05.0, 14.09.01.0) were removed because of their identified
variability. All 12 spectroscopic binaries were double-checked with the
Batten catalogue (1989) and also removed. Since all stars that are
stated as Variable (14.09.00.0) by Simbad were investigated by Cohen
and declared as non variable (with a probability over 90%), their supposed
variability is likely to be due to instrumental uncertainties, so those
stars were not discarded for this reason only. However, because of Simbad
classification rules, a Variable star or a High proper motion star may also
be in a double or multiple system. That is why all stars in Cohen's list
have been searched for in the Catalogue of Visual Double Stars in Hipparcos
(Dommanget & Nys 2000). We have then applied the selection
criteria previously established to all binary systems found. This procedure
led us to reject all Double or multiple stars (12.13.00.0), 4 stars
in double system (14.04.00.0), 3 variable stars (14.09.00.0) and 3 stars
(14.00.00.0). For all stars in double system (14.04.00.0) and stars of other
categories that revealed themselves to be multiple with components within
,
our catalogue provides the separation and the difference in
visual magnitude of the companions (in the field COMTS, see
Table 5). The 3 stars in a cluster (14.01.00.0) have been
kept because each of them happen to be alone in a
field. As a
result, our interferometric catalogue is left with 374 entries.
Let us now discuss some general properties of our catalogue.
Figure 1 displays its sky coverage. The maximum distance
from a point on the sky to the closest calibrator star is
,
whereas the median distance is
.
Most stars (91%) are class III
giants (see histogram in Fig. 2) and all have a spectral
type K (82%) or M0 (18%) (see histogram in Fig. 3). The
majority of stars (72%) have a visual magnitude between 4 and 6, and
almost all of them (95%) between 3 and 7, with a median value of 5.0 (see
histogram in Fig. 4). As for K magnitude, most stars
(95%) lie in the interval K=0-3 with a median value of 1.8 (see histogram
in Fig. 5). Limb-darkened angular diameters (see
Sect. 5.2 below) range from 1 to 10 mas with a median value of
2.3 mas (see histogram in Fig. 6). The median uncertainty
on the diameter is 1.2% (see histogram in Fig. 7) for LD
and UD diameters.
![]() |
Figure 1: Sky coverage of reference stars in the catalogue (Hammer-Aitoff equal-area projection). |
| Open with DEXTER | |
![]() |
Figure 2: Distribution of stars in the catalogue as a function of their luminosity class. Almost all stars (91%) are class III giants. |
| Open with DEXTER | |
![]() |
Figure 3: Distribution of stars in the catalogue as a function of their spectral type. The sample is dominated by K2-K3 (32%), K5 (25%) and M0 (18%). |
| Open with DEXTER | |
![]() |
Figure 4: Distribution of stars in the catalogue as a function of their V magnitude. |
| Open with DEXTER | |
![]() |
Figure 5: Distribution of stars in the catalogue as a function of their K magnitude. |
| Open with DEXTER | |
![]() |
Figure 6: Distribution of stars in the catalogue as a function of their limb-darkened diameter in milli-arcsec. |
| Open with DEXTER | |
![]() |
Figure 7: Distribution of stars in the catalogue as a function of the uncertainty (in percentage) on their limb-darkened diameter. |
| Open with DEXTER | |
The key issue is the maximum interferometric baseline on which
a star of our catalogue can be used as a calibrator. Let us take
the example of a typical star from our catalogue. If modeled as a
uniform disc (UD) in the K band, it would have an angular diameter
mas known with a typical accuracy of 1.2%
(median values). The expected visibility for such a diameter is given by
Eq. (2) and plotted in Fig. 8. The effect of the
uncertainty on the diameter of the calibrator star on the
uncertainty on the visibility can be mathematically derived:
![]() |
(3) |
| Baseline (m) | ||||
|
|
35 | 100 | 150 | 200 |
| 1.2% | 0.1% | 0.8% | 2.3% | 6.4% |
| 5% | 0.4% | 3.5% | 9.4% | 27% |
![]() |
Figure 8:
Visibility curve for a uniform disc model as a function of the
reduced variable
|
| Open with DEXTER | |
It should be pointed out that the visibility drops to zero for x0=3.83 (or
for
in the K band), and that beyond this value, i.e. in the second
visibility lobe and further, limb-darkening and possibly energy at high
spatial frequencies make the uniform disc model a poor approximation of the
real spatial intensity distribution on the stellar disc. Therefore, in order
to work with larger baselines, one has either to refine the reference source
model or to go for smaller and hence fainter objects.
The same issue can be tackled from a different viewpoint that links the
required precision on the visibility to an "effective size'' of the
catalogue. Figure 9 represents as a function of the
baseline, the number of stars whose relative precision on the UD diameter
in the K band is such that they are suitable for the required
labeled on each curve (Table 3 is a numerical
equivalent). In order to ensure easy measurements, we have
retained only the stars whose visibilities would be above 20%.
For instance, if one tolerates a 2% precision on V, then
305 stars in the catalogue can be used as calibrators on a 100 m baseline.
However, only 24 stars remain if a precision of 0.5% is needed.
|
|
||||
| Baseline (m) | 0.5% | 1% | 2% | 5% |
| 50 | 316 | 354 | 366 | 372 |
| 100 | 24 | 186 | 305 | 341 |
| 150 | 0 | 4 | 126 | 266 |
| 200 | 0 | 0 | 4 | 127 |
![]() |
Figure 9:
Number of stars whose formal error on the UD diameter in the K band
is such that they are suitable for the required
|
| Open with DEXTER | |
The catalogue is meant to group together all useful information about calibrator stars in a handy set. We have completed the original fields in C.X: HD number, spectral type, angular diameters and uncertainties on diameters with the equatorial coordinates, HR numbers, common names, parallaxes, proper motions, effective temperatures, surface gravities and Johnson's photometry B, V and from J to N. Two additional quantities useful in the FLUOR reduction process and defined in Sect. 5.3, namely the effective wavenumber and the shape factor, were computed in the case of an ideal interferometer and are provided for the K' band. Measurements have been compiled from different sources or estimated when missing (see Table 5 for details and references). Stars in the catalogue are ordered according to their HD number which we use as an identification number. The reasons why it is essential to keep track of the calibrator stars are twofold:
In their work, C.X derive stellar angular diameters with formal
uncertainties by fitting Kurucz atmosphere models to composite spectra. As
a consequence, such a diameter represents the true physical diameter of the
star in the sense of the Stefan-Boltzmann law. It is not the
uniform disc (UD) diameters that interferometrists are used to, but the UD
diameter can be deduced from this true or limb-darkened (LD) diameter for a
given spectral band provided the limb-darkening coefficients are known. We
have computed UD diameters for J, H, and K bands from the linear
limb-darkening coefficients published by Claret et al. (1995) and
the formula
![]() |
(4) |
| |
Figure 10: Uniform disc (UD) diameter computation schema. |
| Open with DEXTER | |
Equation (2) defines a monochromatic visibility depending on
wavelength for two reasons:
appears obviously in (2),
and also the star's diameter may be a function of
as different
materials are observed at different wavelengths.
As a consequence, it is necessary to define an effective wavelength or
wavenumber for large bands observations (Perrin 1996).
Interferometrists usually prefer to work with the wavenumber
since
it is the Fourier-conjugate variable of the optical path difference.
If we admit that the squared visibility varies slowly with
in the
spectral band of interest, then a first order Taylor expansion yields
![]() |
(5) |
![]() |
|
=![]() |
(6) |
![]() |
(7) |
![]() |
(8) |
![]() |
(9) |
| Sp |
|
| K0-K0.5 | 13.14 |
| K1-K2 | 13.19 |
| K2.5-K3.5 | 13.15 |
| K4.5-K8 | 13.21 |
| M0-M1 | 13.22 |
![]() |
Figure 11: K' band filter transmission profile. |
| Open with DEXTER | |
| Code | Units | Reference(s) | Description | |
| 1 | HD | C.X | Henry Draper catalogue number | |
| 2 | HR | C.X | Bright Star catalogue number, -1 otherwise | |
| 3 | NAME | C.X | Bayer and/or Flamsteed name, HR or HD otherwise | |
| 4 | RA | HH.MMSSmmm | HIP | Right Ascension for Epoch 2000.0 |
| 5 | DEC | sDD.PPSSmmm | HIP | Declination for Epoch 2000.0 |
| 6 | PMRA | mas yr-1 | HIP | Proper motion along RA |
| 7 | PMDEC | mas yr-1 | HIP | Proper motion along DEC |
| 8 | PAR | mas | HIP | Parallax |
| 9 | PARERR | mas | HIP | Uncertainty on PAR |
| 10 | SP | C.X | Spectral type and luminosity class | |
| 11 | LDD | mas | C.X | Limb-darkened disc diameter |
| 12 | LDDERR | mas | C.X | Uncertainty on LDD |
| 13 | TEFF | K | JN87 | Effective temperature |
| 14 | GRAV | dimensionless | SK81 | Surface gravity |
| 15 | LDCJ | dimensionless | C95 | Linear limb-darkening coefficient for the J band |
| 16 | UDDJ | mas | HB74 | Uniform disc diameter in the J band |
| 17 | UDDJERR | mas | HB74 | Uncertainty on UDDJ |
| 18 | LDCH | dimensionless | C95 | Linear limb-darkening coefficient for the H band |
| 19 | UDDH | mas | HB74 | Uniform disc diameter in the H band |
| 20 | UDDHERR | mas | HB74 | Uncertainty on UDDH |
| 21 | LDCK | dimensionless | C95 | Linear limb-darkening coefficient for the K band |
| 22 | UDDK | mas | HB74 | Uniform disc diameter in the K band |
| 23 | UDDKERR | mas | HB74 | Uncertainty on UDDK |
| 24 | EFFWNK | cm-1 | C.X | Effective wavenumber for the K' band |
| 25 | SFK | C.X | Shape factor for the K' band | |
| 26 | B | mag | Simbad | Johnson's B magnitude |
| 27 | V | mag | Simbad | Johnson's V magnitude |
| 28 | J | mag | C.X or AQ | Johnson's J magnitude |
| 29 | FJ | boolean | Flag on J value | |
| 30 | H | mag | C.X or AQ | Johnson's H magnitude |
| 31 | FH | boolean | Flag on H value | |
| 32 | K | mag | C.X or AQ | Johnson's K magnitude |
| 33 | FK | boolean | Flag on K value | |
| 34 | L | mag | C.X or AQ | Johnson's L magnitude |
| 35 | FL | boolean | Flag on L value | |
| 36 | M | mag | C.X or AQ | Johnson's M magnitude |
| 37 | FM | boolean | Flag on M value | |
| 38 | N | mag | J66 | Johnson's N magnitude |
| 39 | FN | boolean | Flag on N value | |
| 40 | COMTS | Simbad, VDS | Comments on the nature of the source, separations | |
| and differences in magnitude for multiple stars |
We have used the CHARM catalogue (Richichi & Percheron 2002) to
quickly search the literature for existing diameter estimates or
measurements concerning stars in our selection. Different methods or
instruments must be analyzed separately in order to reveal systematic
effects at a precision of a few percent. We have chosen references for which
the number of stars is high enough to allow some statistical analysis and
the relative precision is below 10% (Table 6). As a
result, some limb-darkened diameters
can be compared
with photometric estimates given by the infrared flux method: two sets of
respectively 24 stars (Bell & Gustafsson 1989) and 10 stars
(Blackwell & Lynas-Gray 1994) with 5 stars in common, and
direct measurements performed by two interferometers: the Mark III (9 stars)
and the NPOI (21 stars). Some UD diameters in the K band, denoted
,
can be compared as well with direct measurements
performed by two other interferometers: the IOTA (7 stars) and the PTI (10 stars). The six data sets will be investigated in this section.
| Method | Instrument | Compared | Average | Number | Reference(s) |
| diameter | precision | of stars | |||
| IRFM | - |
|
4.0% | 24 | BG89 |
| IRFM | - |
|
4.0% | 10 | BL94 |
| LBSI | Mark III |
|
1.6% | 9 | NS01 |
| LBSI | NPOI |
|
1.9% | 21 | NG99, NS01 |
| LBSI | IOTA |
|
8.4% | 7 | DV98 |
| LBSI | PTI |
|
2.8% | 10 | VB99 |
Let us follow the analysis of Nordgren et al. (2001) for every
data set. In order to look for biases, the mean
and its uncertainty
are computed (
denotes the
standard deviation of the distribution
and N is the number
of stars). Then, the reduced chi-square
is
evaluated in order to check the consistency of error bars. Finally, a linear
least-square fit to the data is performed. If
,
a perfect match would be a=1 and b=0 (Table 7, Figs. 12-14).
| Determination | IRFM (BG89) | IRFM (BL94) | Mark III | NPOI | IOTA | PTI |
| N | 24 | 10 | 9 | 21 | 7 | 10 |
|
|
|
|
|
|
|
|
| 0.10 | 0.05 | 0.16 | 0.13 | 0.64 | 0.17 | |
|
|
0.5 | 0.4 | 2.4 | 3.0 | 1.4 | 6.7 |
| a |
|
|
|
|
|
|
| b |
|
|
|
|
|
|
![]() |
Figure 12:
Comparison with the IRFM: a) angulars diameters determined with
the IRFM vs. values of C.X, b) their differences vs. values of C.X,
c)-d) histograms of the folded distributions
|
| Open with DEXTER | |
The LD diameters measured by interferometry are also in very good
agreement with the values of C.X. Although above one, the
are still very low. Considering the small number of
measures, it is difficult to draw a definitive conclusion about the
compatibility of the error bars. However, the reader may find it appropriate
to increase them by a factor
.
Besides, one may
see a decreasing trend in Fig. 13b, that if confirmed will have
to be explained.
![]() |
Figure 13: Same as Fig. 12 with LD diameters measured by long baseline stellar interferometry. |
| Open with DEXTER | |
Concerning the UD diameters in the K band, the agreement is still good, but there are so few measures that the comparison is merely indicative. As a conclusion, our analysis show that there is no systematic bias between every pair of data sets, and that the error bars are reasonably compatible.
![]() |
Figure 14: Same as Fig. 12 with UD diameters in the K band measured by long baseline stellar interferometry. |
| Open with DEXTER | |
We have presented a catalogue of 374 stars carefully selected to be used as calibrator stars for long baseline stellar interferometry in the near infrared. This catalogue, a subset of the all-sky network of C.X established in the purpose of spectro-photometric calibration, takes advantage of its complete list of angular diameter with formal uncertainties and excellent sky coverage. The comparison between the angular diameters supplied in this catalogue and the measures published by various authors show an excellent agreement. Our catalogue provides additional information on the identification and position of the stars, as well as the Johnson photometry and other useful parameters. Depending on the required precision on the visibility and the wavelength, it is well suited for interferometers with maximum baselines between 100 and 200 m.
Although great care has been put in the double selection process of these stars, first by C.X and then by us, the observational verification of this catalogue has hardly begun. In particular, undetected close companions may be discovered when a high level of a precision is attained. Moreover, as baselines larger than 200 m become available, the number of resolved sources for which the uniform disc model might not be accurate enough will increase significantly (see Fig. 9). More work is needed to extend this catalogue with sources suitable for longer baselines, such as the recently commissionned 330 m-baseline of the CHARA array (e.g. McAlister et al. 2000) or in a few years, the 800 m-maximum baseline of the 'OHANA project (e.g. Perrin et al. 2000). At the same time, this work would benefit to high accuracy instruments coming soon in line. In particular, the VLTI operated with AMBER (Petrov et al. 2000) is expected to attain a 0.1% relative visibility accuracy, necessary to carry out ambitious programs like the direct detection of extrasolar planets.
Acknowledgements
The authors would like to thank Dr. Steve Ridgway for useful discussions and the referee, Dr. Andrea Richichi, for his constructive review of this paper. This research has made use of the Simbad database, operated at the Centre de Données Astronomiques de Strasbourg (CDS), and of NASA's Astrophysics Data System Bibliographic Services (ADS).