A&A 392, 287-294 (2002)
DOI: 10.1051/0004-6361:20020845
C. Grimani1, -
S. A. Stephens7 - F. S. Cafagna2 -
G. Basini6 - R. Bellotti2 - M. T. Brunetti1 -
M. Circella2 - A. Codino1 - C. De Marzo2 -
M. P. De Pascale3 -
N. Finetti1,
-
R. L. Golden4,
-
M. Hof5 -
W. Menn5 - J. W. Mitchell7 - A. Morselli3 -
J. F. Ormes7 - P. Papini8 - C. Pfeifer5 -
S. Piccardi8 - P. Picozza3 - M. Ricci6 -
M. Simon5 - P. Spillantini8 -
S. J. Stochaj4 - R. E. Streitmatter7
1 - INFN and Dipartimento di Fisica dell'Università di
Perugia, Via Pascoli, Perugia 06100, Italy
2 -
INFN and Dipartimento di Fisica dell'Università di Bari,
Via Amendola, Bari 70126, Italy
3 -
INFN and Dipartimento di Fisica dell'Università di Roma II,
Via della Ricerca Scientifica, Roma 00133, Italy
4 -
Particle Astrophysics Laboratory, New Mexico State
University, Box 3PAL, Las Cruces, New Mexico 88003, USA
5 -
Universität Siegen, Siegen 57068, Germany
6 -
Laboratori Nazionali di Frascati,Via E. Fermi,
Frascati 00044, Italy
7 -
NASA/Goddard Space Flight Center, Code 661,
Greenbelt, MD 20771, USA
8 -
INFN and Dipartimento di Fisica dell'Università di Firenze,
Largo E. Fermi, Firenze 50125, Italy
Received 17 January 2002 / Accepted 13 May 2002
Abstract
A measurement of the energy spectra of cosmic-ray positrons and electrons
was made with a balloon-borne magnet-spectrometer, which was flown at a mean
geomagnetic cut-off of 4.5 GV/c. The observed positron flux in
the energy range 7-16 GeV is approximately an order of magnitude lower than
that of electrons, as measured in other experiments at various energies.
The power law spectral index of the observed differential energy spectrum
of electrons is
in the energy interval 7.5-47 GeV. For
positrons the overall fit of the
available data above 7 GeV has been considered. The spectral index is
found to be
and the fraction
of positrons,
/(
+
), has a mean value of
.
The world data on
/(
+
)
from 0.1 to 30 GeV indicate that a plerion
type electron spectrum is preferred over the other types. The trend of the
presently existing high energy data also suggests a possible contribution
of positrons produced at the pulsar polar cap. High resolution experiments
capable of identifying positrons at least up to 100 GeV with high statistics
are required to pinpoint the origin of both electrons and positrons in the
cosmic radiation.
Key words: ISM: cosmic rays
Since the first observation of cosmic-ray positrons in the early sixties (De Shong et al. 1964), many experiments using balloon-borne magnetic
spectrometers have been carried out. It is generally
believed that charged pions produced in the high energy collisions of
cosmic-ray nucleons with the interstellar gas are the dominant source of
via the decay chain
and
.
Based on this
hypothesis, the expected fraction of
in the framework of the Leaky
Box Model appears to be consistent with the observations from 1 GeV to
about 10 GeV. However, above 10 GeV, some measurements indicate an excess
of
with respect to the standard predictions (see the summary in
Golden et al. 1996). In order to explain these observations many
speculations have been put forward. They include
from the radioactive
decay of 56Co and accelerated in supernova remnants (Skibo & Ramaty
1993), by the pair production in the magnetic fields of pulsars
(Grimani 1996 and references therein), as end products
of the annihilation of supersymmetric particles
that were created at the beginning of the Universe (e.g. Tylka 1989)
and from the
interaction of high energy gamma rays with the ambient photons
(e.g. Mastichiadis et al. 1991). It may be noted that except
in the first case, equal amounts of
and
are produced
in all the above processes. Although some recent experiments
(Golden et al. 1996; Barwick et al. 1997;
Coutu et al. 2001)
do not support a large excess of
above 10 GeV,
additional measurements of the energy spectrum of
are essential either
to look for or to rule out these conjectured positron sources.
The large flux of
with respect to
in
the primary cosmic radiation clearly indicates that most of the cosmic-ray
electrons are not of secondary origin, but are produced in cosmic-ray
sources. An accurate measurement of the energy spectra of both
and
would provide an understanding of the spectral shape and the source of
origin of the primary
.
While
can be identified by different
experimental techniques, a magnet spectrometer is essential to distinguish
between
and
.
Many magnet spectrometer experiments have been
carried out to study the electron component, but most of them measured the
positron-to-electron ratio (Boezio et al. 1999 and references therein).
Only 8 measurements in various energy intervals (Fanselow et al. 1969;
Buffington et al. 1975; Golden et al. 1987;
Golden et al. 1994;
Barbiellini et al. 1997; Barwick et al. 1998; Boezio et al.
2000;
Alcaraz et al. 2000) have
been reported on the absolute spectrum of
since 1969. In this paper, a
new measurement of the primary
flux between 7 and 16 GeV, and
electrons between 7 and 47 GeV is presented. The data were collected by the
balloon-borne instrument MASS-91 (Matter Antimatter Space Spectrometer)
flown on September 23, 1991 from Fort Sumner, New Mexico.
The results presented here are compared to the world data
and to recent theoretical predictions.
A schematic diagram of the instrument used in this experiment is shown in Fig. 1. It consisted of a superconducting magnet spectrometer, a time-of-flight device (TOF), a gas Cherenkov counter and a streamer tube imaging calorimeter.
![]() |
Figure 1: Schematic view of the MASS-91 apparatus used in this measurement. |
Open with DEXTER |
The magnet spectrometer consisted of a single coil superconducting magnet
and a hybrid tracking system. The magnet was operated with a current of
120 A producing an inhomogeneous field with a maximum field strength in the
region of tracking device of 2.2 Tesla, decreasing to 0.1 Tesla in the outer
corners.
The tracking system consisted of 2 modules of drift chambers (Hof et al.
1994) and 8 planes of multiwire proportional chambers (MWPC)
adding up to a
total height of 110 cm (Golden et al. 1991).
Each drift chamber (DC) module
consisted of 6 layers
of hexagonal drift cells in the bending direction and 4 layers in the
non-bending direction. The DC system was operated with
.
The spatial resolution of the DC depended on the drift path.
In this experiment,
we achieved an average resolution better than 100
.
The MWPCs
were filled with magic gas and read out by a cathode coupled delay line
system. Three MWPC layers were placed at the bottom of the tracking system,
three were located between the two DC modules, and two layers were kept at
the top of the tracking device. Four MWPCs were instrumented to read out
both coordinates, and the rest measured positions in the bending direction.
One MWPC chamber was not working during the flight. Both the MWPC and DC
tracking systems together provided
19 position measurements along the direction of maximum curvature and 11
in the perpendicular direction. The spatial resolution of the MWPC varied
between 200
and 400
in the bending direction
and was about 1 mm in the non-bending
direction. Because of the inhomogeneous magnetic field
over the spectrometer volume, the deflection error
distribution was found to peak at
3 c/GV, corresponding to
210 GV/c (Papini 1996). This rigidity value has been assumed as the maximum detectable rigidity (MDR).
The present analysis has been limited to 40 GV/c for electrons and to 14 GV/c for positrons, well below the
spectrometer MDR.
The TOF device was made of two planes of scintillators separated by 2.36 m. The upper one was at the top of the gondola and it consisted of two layers of scintillators, each segmented into 5 paddles of 20 cm width and variable lengths to match the round section of the cylindrical payload structure. Each paddle was viewed by a Hamamatsu R2490-01 phototube at the opposite end in each layer, so that the time difference between the signals from each layer can be made use of to locate the position of the particle passing through the paddle. The bottom plane was located just below the tracking device and had one layer of scintillator segmented into 2 paddles, viewed by phototubes at both ends. The coincidence between signals from the two planes provided the trigger for data acquisition. Signals from each paddle were independently digitized for the TOF measurement to determine the direction and velocity, and for the pulse-height analysis to obtain the ionization loss in the scintillators. With this configuration, a timing resolution of 370 ps was achieved for singly charged particles.
The Cherenkov radiator (Golden et al. 1992) was
located in the top section of the payload,
which also housed the top scintillators. It was a 1 m tall cylinder isolated
from the remaining gondola volume and was filled with Freon-12 at a
pressure of 760 Torr at the ground. During the flight, there was a small
change in the temperature and pressure, which resulted in a mean Cherenkov
threshold of
.
The Cherenkov light was reflected
and focused
by 4 segmented spherical mirrors onto 4 BURLE C83073E phototubes. The
response of the Cherenkov detector was position dependent due to the
physical attributes of the mirror and the height, which reduced the path
length of the particle traversing the radiator at the center. Ground muon
data, taken prior to the flight, were used to create a two dimensional
map of the mirror showing the signal amplitude. The detector performance
was described elsewhere (Hof et al. 1996) and we used this information
as a basis to determine the selection criteria for
and
.
The imaging calorimeter was located below the bottom scintillator.
The calorimeter volume was filled with 50 cm long brass streamer tubes of
cross-dimension
mm2. There were 40 horizontal layers, each
consisting of 64 streamer tubes. The tubes in alternate layers were arranged
perpendicularly to each other in order to obtain a three dimensional view of
an
event passing through the calorimeter. The availability of a total of 2560 cells provided a very good granularity to examine the profile of the cascade
development and interactions in the calorimeter (Codino et al. 1989).
The total depth of the calorimeter was 7.3 radiation lengths, which
corresponds to 0.7 inelastic interaction mean free path for protons.
This instrument was launched on September 23, 1991 at 8 am from Fort Sumner, New Mexico, where the geomagnetic threshold rigidity is about 4.5 GV/c. The balloon reached the float altitude of 31.7 km at 10:12 am and after 9.8 hours the transition of the superconducting magnet occured, resulting in the loss of the magnetic field. The flight was later terminated and the mean atmospheric depth at the float altitude, during the period when the magnet was fully operational, was 5.8 g/cm2.
We selected minimum ionizing singly charged particles for the analysis. The particle charge was determined using the dE/dx information from the TOF paddles. We required that the particle deposited less than 1.8 times the most probable energy loss for minimum ionizing singly-charged particles in both the top and the bottom scintillators. These selected events were further subjected to various selection criteria in order to clearly distinguish the electron component from other minimum ionizing particles, such as protons and muons. These criteria are shown in Table 1 and are described below.
![]() |
In order to obtain a reliable information on the measured curvature, from
which both the rigidity and the sign of charge can be determined, the
following criteria were applied. In each event, there should be at least
11 position measurements out of the possible 19 positions in the bending
direction and at least 6 out of 11 positions available in the non-bending
direction. We also required the normalized
in each view to be
8. The error in the deflection measurement also needed to be
0.03. After applying these criteria, negative curvature events were
separated from the positive curvature events for further analysis to
identify electrons and positrons from these samples.
The major source of background for positrons comes from the interacting
protons. Since positrons are expected to produce Cherenkov radiation in
the plateau region, signal from the gas Cherenkov radiator plays an
important role in discriminating positrons against protons. Therefore, we
imposed a strict requirement of at least 8 photo-electrons (pe) to be recorded
by the Cherenkov detector for .
This criterion was relaxed to
0.5 photo-electron for
only to ensure that the event had traversed
the Cherenkov detector. It may be mentioned that no part of the Cherenkov
detector was removed for the analysis. The
selection being very
strict would eliminate particles passing through certain sections of the
Cherenkov detector, which produced low yield, leading to a lower selection
efficiency. This resultant lower efficiency automatically takes
care of the reduction of the geometrical factor due to the loss of effective
area.
Finally, the calorimeter played a very important role in identifying
electromagnetic cascades initiated by the electron component.
The lateral spread of an electromagnetic shower is about 2 Molière
radii, which is approximately 3.2 cm in our calorimeter. We defined
shower cluster in a calorimeter plane, when at least 2 cells were activated within 5 cells from the particle trajectory. The
particle trajectory in the calorimeter was obtained by extrapolating the
trajectory measured by the spectrometer. We required a minimum of 4 shower
clusters in at least one view. The main source of contamination for
is the atmospheric muons and the above criterion rejected all the muons,
as they do not suffer interactions like hadrons. In the case of
we had
to eliminate showers produced by proton interactions and therefore, we
imposed two additional criteria for selecting
.
The first one was
the requirement that the starting point of the shower should be within
the first two calorimeter planes. This reduced the probability of proton
interaction. The second was based on the measured electromagnetic shower
characteristics, namely, that the opening angle of the cascade should be
25 degrees. This criterion removed a large fraction of proton
interactions, as they would have tracks coming out of the interaction
vertex at large angles.
In principle, an upward going
would appear as a downward moving
in the spectrometer.
In this experiment, the TOF measurements can separate upward going particles
from the downward going ones by more than 30 standard deviations. Indeed this
criterion is
redundant for electrons, as the requirement of an electromagnetic shower
in the calorimeter alone makes the criterion for downward moving particle
a very
strict one. Thus, there is no contamination from albedo particles in the
selected sample.
A 17 kg, 1.2 m long aluminium bar with a 7 kg steel swivel in the center was used to connect the gondola to the suspension line from the balloon. The particle loss by cascading and depletion of energy in this bar cannot be reliably estimated due to its peculiar shape, and hence we had rejected all events which came from the direction of this bar. This was carried out by extrapolating the tracks from the spectrometer to the location of the bar. By this procedure we were able to select a clean sample of events, which passed through similar amount of matter before being detected by the instrument and this procedure resulted in only a small reduction in the geometrical factor of about 10%.
With the strict criteria used for selecting the
using the Cherenkov
counter and the calorimeter, the contamination by proton induced events
was negligible in the sample of
selected in this analysis. In order
to ensure this, selected samples of 500
and 300
events above 1 GV/c
were visually scanned for non-interacting particles, representing either
muons or protons, and for low multiplicity hadronic interactions. None
was found and therefore, an upper limit of 1% at 95% confidence level
was set for the background contamination on the selected
and
events
(see Codino et al. 1997).
This visual check also ensured that no spill-over
protons contaminated the electrons in the highest energy interval chosen
for the analysis.
It can also be noticed from Table 1 that we have applied looser
criteria for
compared to
.
These reflect the different background
levels of the
and
.
The ratio p/
in our energy range
is about 103 requiring a rejection factor for protons against
to be at least a few times 104. In the case of
,
the ratio
/
at the float is about one requiring a rejection factor for
muons against electrons of about 100, which is very easily achieved. Thus
the above selection criteria provided a cleaner identification of the rarer
,
but with a slightly lower efficiency, and generated better statistics
for
.
With the above criteria, we selected for further analysis a total of 552 electrons between 5 and 40 GeV, and 37 positrons between 5 and 14 GeV. The selected events below the geomagnetic cut-off were analysed earlier (Codino et al. 1997).
The major parameters required to determine the flux of
and
are the
detection efficiencies and geometrical factor. The procedures for
evaluating these parameters have been reported earlier (Codino et al.
1997) and we describe them briefly below.
Due to the position dependent response of time-of-flight scintillator
counters to cosmic rays and the slightly higher setting of the
discriminators, the particle trigger was not 100% efficient. An
experiment was set up on the ground before the flight with a separate
scintillator telescope to determine this efficiency and it was found that
the trigger efficiency was
.
In order to determine the
efficiency of detectors we selected samples of events using the
remaining devices and applied the selection criteria of
the detector in question. A sample of 20 000 protons gathered during the
flight, which did not interact in the calorimeter, was used to determine the
scintillator selection efficiency and found to be
.
The efficiency of the spectrometer was determined using a two-step procedure.
The TOF system was first used to select events
with Z = 1 and .
The calorimeter was then used to select
electromagnetic showers, whose extrapolated trajectories passed through the
active volume of the tracking system. These selected events were
then fitted with the tracking algorithm and were subjected to the standard
criteria given in Table 1. The efficiency was
.
In order to get the efficiency resulting from the two selection criteria
used in the Cherenkov counter for
and
,
a sample of
were
selected using the calorimeter. These electron events were visually examined
to make sure that the the sample was pure and they were expected to radiate
in the plateau region. We then applied the two different selection criteria for
the Cherenkov counter obtaining an efficiency of
for
and
for
.
The lower efficiency for the
criterion
is indeed
reflecting the loss of regions in the mirror, which were found to produce
smaller number of photo-electrons.
The calorimeter efficiency to record a single track is 100%. We have
examined 293 electromagnetic showers visually and found that the efficiency was only
when applying the requirement of a minimum of
4 shower clusters in at least one view of the
calorimeter. In the case of
selection criteria, the efficiency was reduced from electron efficiency
only by another 2%.
All these detector selection efficiencies are summarized in Table 2.
The geometrical factor was calculated
by analytical integration using the actual magnetic field configuration
and by incorporating the geometrical acceptance employed in the analysis
procedure. We also rejected those parts of the solid angle in which the
trajectories passed through the bar. The calculated geometrical factor
was found to vary below 1 GV/c (Codino et al. 1997). It remained constant
and equal to cm2 sr above this rigidity for both positrons
and electrons. The total flight time used in this analysis is
s, out of which the dead time fraction of the instrument was 0.36.
In Col. 3 of Tables 3 and 4 we show for positrons and electrons,
respectively, the number
of events, which survived the e+ and e- selection criteria described
above.
The two rigidities defining each bin are those measured at the spectrometer. The fluxes at the spectrometer (SPEC) were obtained by dividing the observed number of particles in each rigidity bin by the efficiencies, geometrical factor and total collection time. These differential fluxes are shown in Col. 4. Since the rigidity and energy are nearly the same for these light leptons, the energy interval is identical to the rigidity interval. From a study of the effect of the spectrometer resolution functions (Papini 1996) we inferred that the energy spectrum is not distorted by rigidity measurements in the energy interval of observation of both e+ and e-, since it is well below the MDR. The determination of the energy spectra of e+ and e- at the top of the atmosphere (TOA) requires important corrections to the corresponding spectra observed at the instrument. They include (1) extrapolation of the spectrum to the float altitude after correcting for the Bremsstrahlung and ionization energy losses above the spectrometer, (2) subtraction of the e+ and e- produced in the overlying atmosphere and (3) extrapolation of the resultant spectrum to the top of the atmosphere. These are briefly described below and the details are given in our earlier publications (Golden et al. 1994, 1996; Codino et al. 1997).
The average thickness of material traversed by particles in the detector
before they arrive at the spectrometer is 0.13 radiation length, including
the gondola shell. From the energy spectra at the spectrometer we derived
the spectra of both e+ and e- at the top of the payload by applying
the cascade equation describing the energy loss of electrons due to
Bremsstrahlung and ionization; the corresponding shift in the particle
energy has been also determined. These spectra at the float altitude contain
both the energy degraded primary particles and the secondary e+ and e-
produced in the overlying atmosphere of 5.8 g/cm.
The contribution
of secondary e+ and e- at the float altitude was estimated from
Stephens (1981) and subtracted from the measured spectra at float.
Rigidity | Energy | Number | Flux | Flux |
(GV/c) | (GeV) | of | (m2 sr s GeV)-1 | (m2 sr s GeV)-1 |
TOA | events | SPEC | TOA | |
5-6 | 7.51 | 9 | 0.038 ![]() |
0.020 ![]() |
6-8 | 9.48 | 12 | 0.025 ![]() |
0.014 ![]() |
8-10 | 12.24 | 9 | 0.019 ![]() |
0.011 ![]() |
10-14 | 15.98 | 7 | 0.007 ![]() |
0.0038 ![]() |
Rigidity | Energy | Number | Flux | Flux |
(GV/c) | (GeV) | of | (m2 sr s GeV)-1 | (m2 sr s GeV)-1 |
TOA | events | SPEC | TOA | |
5-6 | 7.51 | 157 | 0.567 ![]() |
0.404 ![]() |
6-7 | 8.88 | 85 | 0.307 ![]() |
0.207 ![]() |
7-8 | 10.24 | 73 | 0.261 ![]() |
0.169 ![]() |
8-9 | 11.61 | 47 | 0.168 ![]() |
0.106 ![]() |
9-10 | 12.98 | 34 | 0.122 ![]() |
0.077 ![]() |
10-12 | 14.95 | 48 | 0.085 ![]() |
0.054 ![]() |
12-14 | 17.71 | 36 | 0.064 ![]() |
0.040 ![]() |
14-16 | 20.46 | 18 | 0.032 ![]() |
0.020 ![]() |
16-20 | 24.34 | 19 | 0.017 ![]() |
0.011 ![]() |
20-25 | 30.42 | 16 | 0.011 ![]() |
0.0071 ![]() |
25-30 | 37.34 | 10 | 0.0071 ![]() |
0.0045 ![]() |
30-40 | 46.92 | 9 | 0.0032 ![]() |
0.0020 ![]() |
The residual spectra thus obtained at float
contain not only the energy degraded primary electron component, but also
a few secondaries produced by the electromagnetic cascading process, whose
contribution to the
is non-trivial. Therefore, the extrapolation
of the residual spectra to the top of the atmosphere was carried out by
simultaneously solving the cascade equations for e+, e- and
-rays, by including Bremsstrahlung, pair production and ionization
processes. The energy shift resulting from the energy loss was used
to derive the flux values of
and
at the top of the
atmosphere. These flux values are given in Col. 5 of Tables 3 and 4. The
median energy in each bin at the top of the atmosphere is also shown in
Col. 2 of Tables 3 and 4. The fraction of positrons in the cosmic
radiation,
,
is shown in Table 5.
The differential flux values of electrons measured by this experiment at the top of the atmosphere are plotted in Fig. 2 by solid circles in the energy region from 7 to 47 GeV.
Energy |
![]() |
(GeV) | |
7.51 | 0.048 ![]() |
9.48 | 0.070 ![]() |
12.24 | 0.107 ![]() |
15.98 | 0.076 ![]() |
It can be seen from this figure that the
energy spectrum can be fitted by a simple power-law in the above energy
domain with a spectral index
.
We have also plotted in this figure the electron spectra measured by magnet spectrometers during the last two decades. There is a general agreement among all these experiments in the energy region covered by this experiment. Notice that the flux values of Golden et al. (1984), shown as open squares, were given for the interstellar space (ISM) and hence the spectrum appears to be steeper than the rest of the data. The present results are consistent with the results from an earlier flight with the same apparatus (except the tracking system), which are shown as open circles in this figure (Golden et al. 1994). This experiment was carried out from Prince Albert, Saskatchewan (Canada) in 1989 at a lower rigidity cut-off of 0.650 GV/c, which extends the spectrum to lower energies. It can also be seen that the AMS spectrum (Alcaraz et al. 2000) seems to deviate from other results below 2 GeV.
![]() |
Figure 2:
Electron energy spectrum measured by MASS-91 (solid circles).
Results from other experiments published in the last twenty years
are also shown. Data have been compared to theoretical calculations
in the interstellar medium (dotted curve: Moskalenko & Strong 1998;
dashed curve: Stephens 2001a,b) and at the Earth by considering the
effect of the solar modulation using a modulation parameter
![]() |
Open with DEXTER |
For a comparison, we have included in Fig. 2 the recent theorethical
calculations of the expected electron spectrum by Stephens (2001a,b),
who used a plerion type injection spectrum. The spectrum shown by the
dashed
curve is for the ISM and that shown by the solid curve is at the Earth with
a solar modulation parameter
GV, which was used by him to
match the results of Boezio et al. (2000). The dotted curve is the calculated
spectrum by Moskalenko & Strong (M&S 1998) in
the ISM and we have modulated
their spectrum using the spherically symmetric model in the force-field
approximation (Gleeson & Axford 1968) with
the same modulation parameter
as in the case of Stephens. We have used the M&S results
based on the propagation without re-acceleration.
Despite different assumptions made by these authors both in the nature of
the injection spectrum and the propagation parameters, there is a good
agreement between these two calculations at energies above 1 GeV.
These calculations also agree remarkably well with the experimental data.
The apparent difference exhibited at lower energies can be tested in the
future with space experiments.
The positron flux values obtained in this experiment are shown in Fig. 3
by solid circles. Symbols used in this and the following figures are the
same as in Fig. 2.
![]() |
Figure 3: Positron energy spectrum measured by MASS-91 (solid circles) and previous experiments. Theoretical curves have the same meaning as in Fig. 2. |
Open with DEXTER |
We have also shown in this figure the theoretical calculations of Stephens as well as those of Moskalenko and Strong; both have estimated the equilibrium spectrum of positrons in the Galaxy resulting from the interactions of cosmic-ray nucleons with the interstellar gas. As in the case of Stephens, we have chosen the estimate of positron spectrum by M&S without re-acceleration and produced by cosmic-ray nuclei having an asymptotic spectral shape as that observed near the Earth. It can be seen that the estimated spectrum of M&S is steeper than that of Stephens. As a result, while Stephens' estimate is in good agreement with the data, M&S estimate falls at the lower end of the data envelope above a few GeV.
The positron ratio defined as the fraction of positrons in the electron
component,
is shown in Fig. 4
along with other existing data obtained during the last two decades
using magnet spectrometers.
![]() |
Figure 4:
Positron fraction versus energy. Solid circles represent
the MASS-91 measurements. Data published in the last twenty years
has also been reported.
Theoretical curves have been estimated from Moskalenko & Strong (1998)
(dotted curve in the interstellar medium, dot-dashed curve near the Earth
by considering a modulation parameter ![]() ![]() |
Open with DEXTER |
Theoretical estimates of the ratio are shown in this figure for comparison. Among these, the calculated ratio by Stephens (solid curve at Earth) is in good agreement with the data especially below 3 GeV and consistent with the general trend in the high energy data within the uncertainties of the measurements. However, the M&S calculations (thin dash-dotted curve) do not fit the data over the entire energy region. At the same time, it is also clear from the figure that the ratio should decrease with energy, based on the secondary hypothesis, specially at high energies. If the ratio either becomes constant or increases above a few GeV, it is essential to invoke the electron-positron pair production process at the pulsar polar cap (Grimani 1996) along with the secondary production in the ISM. The heavy dash-dotted curve in this figure is an example shown by combining the estimated pulsar contribution, assuming a pulsar birthrate of one in 60 years, from Grimani with the M&S curve. Any increase of the ratio can also be explained by the annihilation of supersymmetric particles, but this enhancement would vanish at energies above the mass of such particles (Grimani 2000). It is clear that each one of the above hypotheses has a distinct prediction at high energies, which can be easily tested when measurements could extend to energies of at least 100 GeV.
Acknowledgements
This work was supported by NASA, Grant NAG-110, the National Scientific Balloon Facility, USA, the Istituto Nazionale di Fisica Nucleare (INFN), the Agenzia Spaziale Italiana, Italy and the German Space Agency, DARA, Germany.We thank Mirko Boezio of the Istituto Nazionale di Fisica Nucleare, Trieste, Italy for useful communications.