![]() |
Figure 1: Light curve of a typical sky field observation. Events are extracted from the in FOV region in the energy range 0.5-12 keV; time bin is 30 s. The second part of the observation is affected by intense soft proton flares, the peak count rate is more than 200 times higher than the quiescent one. The selected threshold, in units of sigma of the quiescent count rate distribution (see text and Fig. 2), is marked with a dashed line. |
| Open with DEXTER | |
![]() |
Figure 2: Histogram of the count rate distribution for a typical sky field observation. The corresponding light curve is shown in Fig. 1. The peak corresponds to the quiescent count rate Poisson distribution; points falling to the right correspond to the soft proton flares. The GTI threshold, in units of sigma of the quiescent count rate distribution, is marked with a dashed line. |
| Open with DEXTER | |
![]() |
Figure 3: The cosmic X-ray background spectrum in the 2-8 keV range is displayed, folded with the instrumental response. MOS1 data are represented in grey, MOS2 in black. The best fit model is overplotted. The lower panel shows the residuals in units of sigma. |
| Open with DEXTER | |
![]() |
Figure 4:
The best fit photon index of the CXB is
plotted as a function of the selected GTI threshold (in units of sigma of the
quiescent count rate distribution - see Sect. 3.2). Errors
(at 90% confidence level for a single interesting parameter) include the
uncertainty on the renormalization of the quiescent NXB spectrum. The adopted
3.3 |
| Open with DEXTER | |
![]() |
Figure 5:
Same as Fig. 4, the best fit
CXB normalization is shown. Units are photons cm-2 s-1 sr-1 keV-1 at 3 keV. The contributions from out-of-field scattered light have
not been subtracted. Errors (at 90% confidence level for a single interesting
parameter) include the uncertainty on the renormalization
of the quiescent NXB spectrum. The adopted GTI threshold is marked by the
vertical dashed line. It is evident that no significant variations are
obtained as a result of small changes of the selected GTI threshold (within
the range |
| Open with DEXTER | |
![]() |
Figure 6:
Best fit CXB photon index as a function of
the maximum allowed ratio for the 8-12 keV surface brightness
|
| Open with DEXTER | |
![]() |
Figure 7:
Same as Fig. 6, the best
fit CXB normalization (in units of photons cm-2 s-1 sr-1 keV-1 at 3 keV) is plotted here. The stray light correction have not been
applied. Errors (at 90% confidence level for a single interesting parameter)
include the uncertainty on the renormalization of the
quiescent NXB spectrum. When dataset having large values of
the ratio
|
| Open with DEXTER | |
![]() |
Figure 8: Best fit CXB photon index as a function of the off-axis angle. Each point represents an independent measure obtained from the analysis of a selected portion of the detector plane (see text). Errors (at 90% confidence level for a single interesting parameter) include the uncertainty on the renormalization of the quiescent NXB spectrum. The MOS1 case is shown (MOS2 is very similar). The constant trend is a result of the correct vignetting function calibration. |
| Open with DEXTER | |
![]() |
Figure 9: Same as Fig. 8, the case of the CXB normalization is presented here. Units are photons cm-2 s-1 sr-1 at 3 keV. The corrections for the light coming from out-of-field angles have not been applied. Errors (at 90% confidence level for a single interesting parameter) include the uncertainty on the renormalization of the quiescent NXB spectrum. The variations are always smaller than the uncertainty, confirming the overall correctness of the vignetting curve calibration. |
| Open with DEXTER | |
![]() |
Figure 10:
CXB intensity measurements. The flux in the 2-10 keV band is
represented as a function of the epoch of the experiment. The plot is an update of Fig. 3 of
Moretti et al. (2003) including the results of the present work. From left to
right the CXB values are from Marshall et al. (1980) with HEAO-1 data; McCammon
et al. (1983) with a rocket measurement; Gendreau et al. (1995) with ASCA SIS data;
Miyaji et al. (1998) with ASCA GIS;
Ueda et al. (1999) with ASCA GIS and SIS; Vecchi et al. (1999) with BeppoSAX
LECS/MECS; Lumb et al. (2002) with XMM-Newton EPIC/MOS; Kushino et al. (2002)
with ASCA GIS. Finally, the hollow star mark our own CXB measurement with the EPIC MOS cameras onboard XMM-Newton. All the uncertainties are at 1 |
| Open with DEXTER | |
![]() |
Figure A.4: The correlation between the surface brightness (units of counts cm-2 s-1) in the 1.4-1.6 keV range (dominated by the internal Al-K fluorescence line - see Fig. A.1) and the surface brightness in the range 2.5-5 keV (continuum). Data are for closed observations; MOS1 red and MOS2 black. A rather high scatter is evident. |
![]() |
Figure A.5: Same as Fig. A.4; the surface brightness (units of counts cm-2 s-1) in the energy ranges (2.5-5 keV) and (8-12 keV) show a better correlation. |
![]() |
Figure B.2: Same of Fig. B.1 for the case of the closed observations. The correlation is much better. |
![]() |
Figure B.4:
The ratio
|
![]() |
Figure B.9:
Best fit CXB parameters computed from the three independent dataset corresponding to different values of
|