A&A 411, L209-L213 (2003)
DOI: 10.1051/0004-6361:20031225
L. Natalucci1 - A. J. Bird2 - A. Bazzano1 - P. Ubertini1 - J. B. Stephen3 - R. Terrier4 - L. Lerusse5
1 - CNR-Istituto di Astrofisica Spaziale e Fisica Cosmica,
Area Ricerca Roma 2/Tor Vergata, via del Fosso del Cavaliere 100,
00133 Roma, Italy
2 -
Department of Physics & Astronomy, University of Southampton, SO17 1BJ, UK
3 -
CNR-Istituto di Astrofisica Spaziale e Fisica Cosmica,
Area Ricerca di Bologna, via Gobetti 101, 40100 Bologna, Italy
4 -
DAPNIA, Service d'Astrophysique, CEA/Saclay, 91191 Gif-sur-Yvette Cedex, France
5 -
INTEGRAL Science Data Centre, Chemin d'Écogia 16, 1290 Versoix, Switzerland
Received 14 July 2003 / Accepted 8 August 2003
Abstract
The spatial distribution of the background events may
affect the source detection capability of IBIS at high
energies (
200 keV) for both ISGRI and PICsIT layers.
The observed background is found to be variable
and spatially structured, and in some cases its properties
strongly deviate from the expected statistical behaviour.
Background correction methods are then necessary to
improve the quality of the shadowgrams obtained from sources.
In order to perform an efficient flat-fielding the
response of the detector to both source (
-rays) and
background events is investigated using data from Monte
Carlo simulations and in-flight calibration observations.
Key words: gamma-ray astronomy - detectors - background - data analysis
The IBIS instrument on board INTEGRAL (Winkler et al. 2003) is the first
large area, space born
-ray telescope carrying a pixellated,
multi-layer detector system (Ubertini et al. 2003). Due to its large
aperture and large collecting area, the scientific
performances are strongly dependent on background reduction.
Optimal levels of background intensity and
stability in time have been reached after tuning and
calibration of instrumental parameters. This has been achieved
during in-flight commissioning (Ubertini et al. 2003).
In IBIS, different data acquisition modes working simultaneously
ensure a good capability of background rejection. Important features
are the selection of single and multiple events in the high energy
detector (PICsIT, Labanti et al. 2003) and of multi-layer coincidence events
(Compton mode). Furthermore, the two detector layers
ISGRI (the hard X-ray detector, Lebrun et al. 2003) and PICsIT are actively
shielded by a multi-module VETO system, based on one lateral and
one bottom arrays of BGO blocks (Quadrini et al. 2003). ISGRI is also shielded
from the diffuse hard X-ray background outside the field-of-view by a
composite passive shield protecting the whole telescope from mask
to detector base (see e.g. Natalucci & Caroli 1996).
The IBIS data processing software (Goldwurm et al. 2003) is
capable of recovering many time varying systematic effects as
telemetry losses,
noisy pixels, temporary switch-off of ISGRI modules,
dead time etc., which are well quantified and properly taken
into account in the analysis. Once the data have been corrected
for these effects we still see important spatial structures, which
affect both detector layers. These are easily detected as
different average count-rates in
detector modules/semi-modules, enhancements at the edge of the modules
and local effects induced by the readout electronics.
There are several factors known to produce this inhomogeneity of the
spatial response (see also Stephen et al. 2003; Terrier et al. 2003):
a) a spatially dependent hadronic background component, depending on
the external payload and spacecraft mass distribution;
b) the efficiency of detection of single and multiple events,
which is expected to be dependent on the detector position;
c) the individual response of each detector module; d) the intrinsic
response of each pixel as a function of its position, i.e. its
proximity to a module, an ASIC (Application Specific Integrated
Circuit) readout element and/or a dead pixel; e) variations
of the performance of the VETO modules, induced especially by
temperature variations.
These effects are, in turn, energy dependent. Moreover a given pixel
may undergo periods of unusual behaviour. For ISGRI, defective pixels
including noisy elements which are switched off permanently or for a
relatively short time period may induce local variations in the detector
itself and even in PICsIT.
PICsIT pixels have also intrinsic noise levels that could have local
effects in ISGRI. PICsIT data are normally transmitted as
spectral histograms (SI) already accumulated on board.
Varying count rates of pixels can influence the local response of
detectors for less than the typical SI integration time (
2150 s),
and hence cannot be properly followed in PICsIT. For this reason, local
corrections based on analysis of single pixel light curves cannot be
always applied efficiently. Therefore, methods of flattening the images
independently of the time behaviour of pixels are being studied
(see Sect. 2).
The difficulty of correcting data by using maps obtained by long background exposures is mainly related to the background being variable with time, and to different systematics in the counting behaviour of pixels in "observation'' and "background'' exposures. In the following sections, an overview of the main problems related to background induced effects is given and some results are reported in more detail, especially regarding the PICsIT detector. For PICsIT, an important source of background is the excess counts caused by the decay of long-lived phosphorescence states of the CsI crystals (Hurley 1978). These counts cannot be vetoed because they give rise to single events, separated by intervals of the order of a few 10-5 s. However, they can be easily identified and removed in photon by photon data, and their spatial distribution can be separately studied (see Sect. 4). Other spatial inhomogeneity effects which are background independent are described in Sect. 3.
![]() |
Figure 1: Example of PICsIT model map in the energy range 190-300 keV. It shows different average levels associated to the 16 semi-modules, the edge pixels, and pixels surrounding dead pixels (the latter are shown in black). The image axes (in units of pixel index) are parallel to the spacecraft Y- and Z-axes, for which the X-axis represents the pointing direction. |
| Open with DEXTER | |
For IBIS, flat-fielding prior to image deconvolution (Goldwurm et al. 2001)
is essential especially at high energies,
as long as the size of the fluctuations in the background level,
propagated through the imaging process exceed the observed celestial source
intensity. The approach of applying a flat-field based on Fourier filtering
is difficult, because a substantial power of the background distribution
is seen at high spatial frequencies, where most of the signal from
multiplexing the source counts by the coded mask is also found.
An alternative approach is
to characterize each pixel by averaging on pixel categories, defined on the
basis of the pixel position in the detector, i.e. proximity to a module edge
and/or to a noisy/dead pixel, position within an ASIC etc. This
has the form of a model map applicable to a given energy band. The model
is obtained from real data by measuring average count rates for each pixel
category (see Fig. 1).
Correction maps defined for eight standard
energy bands, covering the range 0.2-6.5 MeV, have been used
to subtract a background shape from the single event shadowgrams.
The resulting detector image is defined by a subtraction of the model from
the original image:
,
where M is
the model map and
is the ratio between average
pixel values in the raw image and model map.
The standard deviation in the pixel counts of the background subtracted
shadowgram is compared to that of the original image, and a flat-field
efficiency is evaluated as:
![]() |
(1) |
![]() |
Figure 2: Flat-field efficiency as a function of energy, for the model maps discussed in Sect. 2. |
| Open with DEXTER | |
Adding more input data in building the model maps is not found to improve the efficiency. This is probably due to the time variations in the background. Improvements to the model are currently under study, as long as more systematic effects are identified.
![]() |
Figure 3: Simulated map of ISGRI, as illuminated by a monochromatic source at 500 keV. The energy range is 170-400 keV. The image shows a non perfectly flat events distribution. Both axis scales refer to pixel index (0 to 127 on each side). Axes are parallel to the spacecraft Y- and Z-axes. |
| Open with DEXTER | |
Besides the known effect of background non-uniformity, scattering of
celestial source
-rays associated with a non-ideal response of the
event selection logic and/or non-ideal geometry of the detector can produce
a non-homogeneity of the source count distribution. This component of the
non-uniformity (which is intrinsically different from the background
induced one) is being studied by Monte Carlo simulations using
a numerical code based on a detailed geometric and mass model of the telescope,
including electronics and event selection logic (Laurent et al. 2001).
For this purpose, we have simulated a
parallel flux of
-rays incident on the IBIS detector with zero
inclination, at different monochromatic photon energies and for a power
law source. The efficiency and response of single pixels are considered
uniform in the simulation, so inhomogeneities cannot be ascribed to their
local behaviour. For a monochromatic source, it was verified that the image is
perfectly uniform when selecting events under the full-energy peak. This
is expected as far as there is no re-distribution of the event energy in
pixels other than the incidence pixel. In contrast, the count distribution
shows differences,
especially in the counting rate at the edge of modules,
when the energy range selected covers part of the Compton continuum.
This is true in
particular for ISGRI (see Fig. 3) and PICsIT multiple events. This
non-uniformity should be ideally corrected by dividing the residual detector
image after background subtraction or flat-field (see Sect. 2) by
an efficiency map
which takes into account the scattering induced effects. This is, of
course, dependent on the input source spectrum. A set of reference
efficiency maps are being computed for a limited set of input
spectra models (based on power law or cutoff power laws) to be tested
and compared. We consider, however, that
correcting real data using a non-uniformity model of the source
illumination is expected to be of secondary importance compared to
an efficient background subtraction.
![]() |
Figure 4:
Light curves of a sample of photon by photon data for
PICsIT single events. Events are rebinned on 100 s.
Plotted errors are the expected Poisson
uncertainties (1 |
| Open with DEXTER | |
The PICsIT detector is found to show significant count-rate increase caused
by fake events, induced by decay of phosphorescent states in the CsI
crystals. These states are excited by passage of particles in
the active body of the PICsIT detector, which occur at a rate
1000
times lower than the normal background rate. One burst
of fake events can be produced by either the passage of a primary particle,
or by a secondary shower (cascade) producing a track or other figure of
well-defined spatial pattern (Segreto et al. 2003). All these counts are mostly
detected at energies below
300 keV and can be easily recognized in
photon by photon data, in which we have full timing information.
Since each individual passage of particles produces a relatively large number
of these events, the count rate statistics is strongly modified.
The two panels of Fig. 4 show the fluctuations of the summed count rate
for 16 pixels of an ASIC, in the energy range 170-285 keV. On the upper plot,
representing the total counts, the fluctuations are clearly non-Poissonian.
Once the phosphorescence triggers are eliminated, the fluctuations are fairly
compatible with Poissonian noise (lower panel). It is quite easy to recognize,
by comparing the two plots, that some time bins are not affected at all
by phosphorescence triggers. This effect can be quantified
by computing the ratio, R, between the observed standard deviation of the
counts and the Poisson statistical error, as a function of energy and
for different integration times. In Fig. 5 (upper panel) are shown
the values of R obtained from the same set of data (16 pixels,
s). From the figure, it is clearly seen that the loss of
sensitivity per channel does not depend on integration time for
10 s.
![]() |
Figure 5:
The ratio of the observed standard deviation to the Poissonian expected
value, as a function of energy bin, for PICsIT photon by photon
data (top) and for PICsIT spectral histograms (bottom). The three curves
in the upper panel refer to different integration times: 10 s (black),
100 s (red), 500 s (green). The integration time of histograms is
fixed to 2150 s. In the upper panel, data are plotted against PHA channels,
whilst in the lower panel, SI energy bins are used. Both X-axis scales are linear
in energy, with a conversion factor |
| Open with DEXTER | |
We have then investigated the effect of these fake events
in spectral imaging
data, which is the PICsIT standard mode of operation, using an empty
field exposure (integration time: 2150 s). As the background intensity
always shows a long-term trend,
expected values of rates were computed by a linear fit to the
data, after selecting an interval of constant linear variation of
25 hours. For this test, we have used total count rates and verified
that a large spreading of the values is also present in the energy
channels below
250 keV (see Fig. 5, lower panel).
The ratio R changes significantly when rebinning in energy. For example,
we estimate
and
for the energy bands
170-220 keV, and 220-280 keV respectively. These values are sensibly
higher than the ones plotted in Fig. 5 for the single energy
channels. This is actually expected, as
each track produced by a particle leaves a number of phosphorescent
counts which are widely distributed in energy, thus giving rise to
highly correlated count rate fluctuations in different energy channels.
Finally, we investigated the spatial distribution of the
fake events by using the photon by photon data.
The images obtained in the energy band 170-285 keV clearly show
that the non-uniformity in this range cannot be ascribed to the
phosphorescence triggers (see Fig. 6).
![]() |
Figure 6: The spatial distribution of normal background events in PICsIT (upper image), compared to that of phosphorescence triggers in the energy band 170-285 keV (lower image). The image axes are the same specified in Fig. 1. |
| Open with DEXTER | |
Acknowledgements
The IBIS project is partially granted by Italian Space Agency (ASI). AJB is funded by PPARC grant GR/2002/00446.